Thermo-shaker PCMT
Version 1.01 - January 2013
Page 10
Operating Instructions
Fig.6 Control panel in calibration and operation mode
5.10. Set the temperature with multiplier coefficient readings (Fig 6A/1) to change the actual
temperature value of the device.
Using the
Temp.
▲ and ▼ keys (Fig.3/3), change the calibration coefficient (Fig.6A/1) so, that
the new temperature value (Fig.6A/2) corresponds to the independent temperature sensor. In
our example the calibration coefficient will be 0.974 (in range: 0.936 up to 1.063; increment 0.001).
Caution!
This calibrating coefficient will correct the temperature through the whole operating
range.
5.11. After finishing the calibration press the
RPM-RUN/STOP
key (Fig.3/4) once to save the
changes and exit the calibration mode.
5.12. The display will show calibrated temperature as shown on fig.6B/3 and the unit will continue
thermal stabilization according to the previously set temperature.