4. The cuff will deflate quickly and the blood pressure reading, pulse
reading will show in the display. Then the icon ‘ ’ will flash to remind
the user to record the reading as shown in Fig 8;
5. Press button ‘M1’ or button ‘M2’ to record the reading in the
corresponding memory; the measurement time will be displayed on the
screen alternately. If the user does not press the button, the reading
won’t be recorded;
6. Press the button ‘ ’ to turn off the device. Rest for at least 5 minutes
before taking another measurement. If the power supply is not
switched off and the device is unused for 3 minutes, the device will
switch off automatically.
AUTOMATIC INFLATION
There are four inflation pressure levels for this device: 190mmHg,
230mmHg, 270mmHg and 300mmHg. When 190mmHg is not enough or
arm movement occurs, the device will automatically inflate to higher
pressure to ensure a successful measurement. It is not a fault.
RAPID DEFLATION DURING MEASUREMENT
If you do not feel well during measurement or want to stop the
measurement for some reason, press the button ‘ ’. The device will
quickly release the air in the cuff and the device will be returned to
standby mode.