Calibration
12
3A1657A
18. Enter 0 in the Alarm % field. This will prevent alarms
that during the calibration dispense that could pre-
vent dispensing.
19. Press
to exit the System screen then use the
arrow keys to navigate to the Calibration screen.
20. Press
to access the weight calibration and
material specific gravity screen.
21. Enter the material specific gravity in the A (Red) and
B (Blue) specific gravity input boxes.
22. Press
to go to the main Calibration screen.
23. Press
to go to the flow meter calibration
screen.
24. Enter a value in the Cal. Setpoint input box that is
close to what is used during normal dispensing.
25. Prepare containers for calibration dispense:
For L-Head, S-Head, and GX-16 Dispense Valves
Only --
a.
Weigh a container and record the weight.
b.
Place container below dispense valve.
For Fusion, EP, Probler P2, and MD2 Dispense
Valves Only --
a.
Mark two containers as A and B.
b.
Weigh each container and record the weights.
c.
Place containers below the dispense valve to
catch the each fluid separately.
All Assemblies --
26. Press
to begin dispensing then press again to
stop.
27. Weigh the container(s) and record the weight(s).
Subtract the weight of each bucket measured before
the shot to obtain the weight of each material dis-
pensed.