5
|
Instructions for Use
|
CultureWell™ MultiWell Chambered Coverslips
11.
To mount coverslips on glass microscope slides, first remove the silicone gasket by peeling it off
as shown in Figure 10.
Figure 10
Specimens may also be mounted using aqueous media by filling wells and pressing another
coverslip onto the gasket.
12.
To Clean and Re-Use MultiWell Gaskets
Rinse off reagents immediately after each use. Wash the gaskets with distilled water and if
necessary with a mild detergent that is suitable for cell culture. Do not soak gaskets in detergent.
Rinse thoroughly with distilled water and allow gaskets to dry. When re-using cleaned gaskets,
just prior to applying them to a new coverslip, use a piece of adhesive tape to clean lint and
debris from the gasket surface.
Place a clean, dry 24mm x 50mm glass coverslip onto the silicone gasket and press gently to seal.
Place gasketed coverslips individually into glass petri dishes (silicone side up), wrap with
aluminum foil and sterilize by autoclaving. Allow coverslips to cool and dry completely before
use.
CultureWell™ MultiWell chambered coverslips are for laboratory use only.