SPECIFICATIONS  –  LOCAL  MODEL

Applications:

Use for indoor or outdoor applications where occasional 

moisture is common.

Materials:

Acetal,  Amorphous  Nylon,  Silicone  Rubber,  Polyester 

(decals), Viton (gasket & seals), Stainless Steel (fasteners)

Power Source:

2-wire, loop powered

*

. 8.5 VDC to 35 VDC loop voltage 

required for correct operation.

Outputs

**

:

Analog Primary:  4-20 mA current loop. With loop voltage 

maintained within specified limits, will maintain advertised

linearity over 4 mA to 20 mA range, with good linearity in 

over-range conditions to about 1.5 mA and 25 mA.
Analog Auxiliary:  0 to 5 VDC. Relationship to primary cur-

rent output: V(volts) = l (ma) ÷ 5 . Will maintain advertised 

linearity over 0.8 V to 4.0 V range, with good linearity in 

over-range conditions to about 0.1 V and 4.9 V.

Recommended minimum driven impedance = 10K Ohms.
Digital  (Pulse-Out):    Open  Collector,  square  wave.  Will 

switch up to 60 V and up to 200 mA. Closed circuit volt-

age drop typically 0.1 V; guaranteed less than 0.5 V at any 

current up to 200 mA.

Configuration:

2-Totals (1 cumulative and 1 batch), Rate, 2 Cals

(Factory calibration in gallons or litres; 1 field calibration)

Input Signal:

Hall Effect,  Reed Switch, NPN, Open Collector or 

Sine Wave

Time Base:

Hours or minutes

Unit of Measure:

U.S. gallons or litres

Accuracy/Performance:

Possible conversion error, in addition to any inaccuracy of 

coupled flowmeter, as follows:
Max. Conversion Error:  (nonlinearity plus span, any input, 

loop current output, 0°C to 70°C, loop voltage supply 12 

VDC to 24 VDC) 0.5% of span plus possible resolution 

uncertainty.
Max. Conversion Resolution Uncertainty:  (Loop current 

output, when properly calibrated) Larger of 0.1 mA or 

[20 mA / (10 x ( f

20

–f

4

)]. Where f

20

 = frequency at 20 mA,

f

= frequency at 4 mA.

Speed of Response:  After step change in input frequency, 

loop output guaranteed stable within 3 x accumulating time 

(Accumulating time user selectable from 0.3 sec, 0.7 sec, 

1.3 sec, 2.6 sec, 5.2 sec).

Frequency Range:

0.25 to 1,000 hertz

Batch Total:

Up to 999,999 (x100)

Cumulative Total:

Up to 999,999 (x100)

Temperature:

  Ambient Temperature:  +32° F to 140° F (0° C to 60° C)

Cable:

No cable provided

Mechanical Connections:

Display is mounted directly to flow meter body.

Electrical Connections:

GX500 - Two strain relief ports

GX510 - One strain relief port: one threaded plug

G2 Series - Two strain relief ports

GM Local Models - Two strain relief ports

GM 1/2 in. and Larger - One strain relief port: one 

threaded plug

Shipping Weight:

1.1 lb. (.5 kg)

*

 

 Models utilizing Reed Switch pickup are completely loop powered, and 

do not require any additional power supply. Note, however, that models 

utilizing Hall Effect signal pickup require an additional independent, iso-
lated DC power supply for powering the Hall Effect device.

**

 

 If you want to use the 4-20 mA output 

and 

one or both of the other 

outputs at the same time, you must provide electrical isolation between 

the current loop and all other circuit elements. Failure to do so will result in 

incorrect 4-20 mA signal conversion, and possible damage to the unit.

Summary of Contents for GM001

Page 1: ...CONTENTS To the owner Congratulations on receiving your GPI 4 20 mA Out with Display We are pleased to provide you with a product designed to give you maximum reliability and efficiency Our business i...

Page 2: ...in indoor or outdoor applications where occasional exposure to moisture is common The 4 20 mA Out with Display is available in two ver sions One indicates flowrate in units minute and one in units ho...

Page 3: ...her liquids occasional low pressure splashing will not harm unit if cable entry points are well sealed subject to 5g shock loading facing the sun directly for long periods of time close to high voltag...

Page 4: ...m ONLY Local Com HL HF IN R See note EDM Power INPUTS OUTPUTS 4 20 mA Out with Display Loop Loop current current Min V 8 5 Max V 35 Customer Equipment Minimum signal amplitude required for normal oper...

Page 5: ...Max V 35 Customer Equipment Loop Power Supply Typ 12 24 VDC Loop Input Loop Input NOTE 1 Loop power supply electrically isolated from customer equipment NOTE 2 Actual value of loop current mA is disre...

Page 6: ...Frequency Generating Input Device INPUTS OUTPUTS 4 20 mA Out with Display Loop Loop current current Customer Equipment Loop Power Supply Typ 12 24 VDC Open Collector Pulse In Pulse In Return Pulse Ou...

Page 7: ...that the enclosure sealisnotcrimpedortwisted Donotover tightencornerscrews hand tighten only Faceplate can be rotated 90 W I R I N G D I A G R A M 8 4 20 mA or 0 20 mA Output Customer Equipment Witho...

Page 8: ...4 20 mA and Pulse Output option available only with Reed Switch input W I R I N G D I A G R A M 1 0 4 20 mA or 0 20 mA Output and Pulse Output Customer Equipment With Built in Power Supply Input Reed...

Page 9: ...button for 3 seconds to reset the Batch Total to zero When fluid is flowing through the meter a small propeller icon is highlighted Flowrate Feature Tousethisfeature pressandreleaseDISPLAYuntilFLOWRA...

Page 10: ...ser to tweak the factory calibration by a percent that represents application fluid or plumbing differences 1 To calibrate press and hold the CALIBRATE and DISPLAY buttonsforabout3secondsuntilyouseeFL...

Page 11: ...the transmitter If signal is being received deactivate it by one of the following methods Stop the flow through the line that is being recorded Disconnect the output device on the flowmeter Locking U...

Page 12: ...presently operating temporarily disable its external power supply Be sure to allow at least 30 seconds to elapse with unit unpowered 2 Press and hold the 4 button While holding watch the indicator li...

Page 13: ...signal input connections for intermittent open or short circuits F LOOP OUTPUT 1 Output response time setting too long Select a shorter response time setting STABLE BUT especially for fast input sign...

Page 14: ...14 DIMENSION DRAWINGS...

Page 15: ...Tapping GM001 GM002 and GM003 2 9 120058 01 Bracket 1 10 904005 13 Screw 6 32 x 1 2 in 4 Item No No Part No Description Req d 11 904002 44 Screw 8 32 x 5 16 in 2 12 125066 20 Cable 20 ft 1 125066 3 C...

Page 16: ...linearity plus span any input loop current output 0 C to 70 C loop voltage supply 12 VDC to 24 VDC 0 5 of span plus possible resolution uncertainty Max Conversion Resolution Uncertainty Loop current o...

Page 17: ...gital Pulse Out Open Collector square wave Will switch up to 60 V and up to 200 mA Closed circuit voltage drop typically 0 1 V guaranteed less than 0 5 V at any in range current Configuration 2 Totals...

Page 18: ...Please contact GPI before returning any part It may be possible to diagnose the trouble and find a solution with a telephone call GPI can also inform you of any special require ments you will need to...

Page 19: ......

Page 20: ...rranty shall not apply if A the product has been altered or modified outside the warrantor s duly appointed representative B the product has been subjected to neglect misuse abuse or damage or has bee...

Reviews: