GPI GM001 Owner'S Manual Download Page 13

13

SYMPTOM 

PROBABLE CAUSE 

CORRECTIVE ACTION

A.  METER IS NOT 

1.  Field Calibration not performed properly.  Field Calibrate again or select Factory Calibration.

 

 

2.  Factory Calibration not suitable 

Perform a Field Calibration according to Calibration Section or

 

 

  for liquid being measured. 

select the proper Factory Calibration selection (i.e., gallon or litre).

 

 

3.  Improper installation of flowmeter. 

Check for electrical noise, pulsation or swirl in the flow.

 

 

4.  Flowrates too high or too low. 

See section on display calibration for flowrates.

B.  READOUT FADED 

1.  Power not connected. 

Check power supply.

 

 

2.  Wiring incorrect. 

Verify connections.

 

 

3.  Computer defective. 

Contact the factory.

 

 

4.  Temperature limits exceeded. 

Check temperature specifications.

C.  NORMAL FLOW- 

1.  Field Calibration not performed correctly.  Field Calibrate again or select Factory Calibration.

 

 

2.  Computer defective. 

Contact the factory.

 

  

3.  Loose wire or mis-wired. 

Check wiring diagram or cable installation.

 

 

4.  Sensor not attached to turbine. 

Check continuity of sensor.

 

 

5.  Faulty sensor. 

Contact the factory.

D.  LOOP OUTPUT 

1.  4 mA / 20 mA setpoints bad or not set. 

Perform new setpoint procedure for both 4 mA and 20 mA points.

 

WITHIN NORMAL 

   

 

RANGE, BUT

 

INCORRECT

E.  LOOP OUTPUT 

1.  Output response-time setting too short, 

Select a longer response-time setting.

 

“BOUNCES” 

  especially for slow input signal.

 

 

2.  Input connections bad. 

Check all signal-input connections for intermittent open- or

 

 

   

short-circuits.

F.  LOOP-OUTPUT 

1.  Output response-time setting too long, 

Select a shorter response-time setting.

 

STABLE BUT 

  especially for fast input signal.

 

RESPONSE TIME 

 

 

TOO SLOW

G.  LOOP-OUTPUT OK, 

1.  0-5 V output loaded too heavily. 

Be sure 0-5 V load impedance is at least 1000 ohms (1K

Ω

).

 

 

2.  Wiring incorrect. 

Verify connections.

H.  LOOP OUTPUT  

1.  Loop not supplying power. 

Be sure loop power supply is present and working, and has 

 

 “STUCK”AT ZERO 

   

correct polarity.

 

 

2.  Loop connections bad. 

Check all loop connections for open- or short-circuits.

 

 

3.  Transmitter is faulty. 

Replace transmitter.

I.  LOOP OUTPUT 

1.  4 mA / 20 mA setpoints bad or 

Perform new setpoint procedure for both 4 mA and 20 mA

 

 “STUCK” AT LOW  

  not set. 

points.

 

VALUE  (Between 1 

 

and 4 mA) RE- 

2.  No input signal. 

Verify presence of input signal at terminal block.

 

GARDLESS OF 

 

INPUT SIGNAL 

3.  Input connections bad. 

Check all signal input connections for open- or short-circuits.

 

 

 

4.  Unit is faulty. 

Replace unit.

J.  LOOP OUTPUT 

1.  4 mA / 20 mA setpoints bad or 

Perform new setpoint procedure for both 4 mA and 20 mA

 

 “STUCK” AT FULL-     not set.  

points.

 

SCALE (above 

 

20 mA) REGARD- 

2.  Short-circuit between Loop (–) 

Check all Loop and LOCAL-COM circuitry for shorts.

 

LESS OF INPUT 

  and LOCAL-COM circuits.

 

SIGNAL 

 

 

3.  Incorrect connection of Hall 

Hall Effect requires dedicated power supply.

 

 

  Effect device.

TROUBLESHOOTING

OR BLANK

ACCURATE

RATE  BUT  METER 

DOES NOT COUNT

(Meter comes on

when DISPLAY

button pushed)

ERRATICALLY

(is unstable)

BUT 0-5 V OUTPUT

DOESN’T WORK

(No reading at all,

regardless of input

signal.)

Summary of Contents for GM001

Page 1: ...CONTENTS To the owner Congratulations on receiving your GPI 4 20 mA Out with Display We are pleased to provide you with a product designed to give you maximum reliability and efficiency Our business i...

Page 2: ...in indoor or outdoor applications where occasional exposure to moisture is common The 4 20 mA Out with Display is available in two ver sions One indicates flowrate in units minute and one in units ho...

Page 3: ...her liquids occasional low pressure splashing will not harm unit if cable entry points are well sealed subject to 5g shock loading facing the sun directly for long periods of time close to high voltag...

Page 4: ...m ONLY Local Com HL HF IN R See note EDM Power INPUTS OUTPUTS 4 20 mA Out with Display Loop Loop current current Min V 8 5 Max V 35 Customer Equipment Minimum signal amplitude required for normal oper...

Page 5: ...Max V 35 Customer Equipment Loop Power Supply Typ 12 24 VDC Loop Input Loop Input NOTE 1 Loop power supply electrically isolated from customer equipment NOTE 2 Actual value of loop current mA is disre...

Page 6: ...Frequency Generating Input Device INPUTS OUTPUTS 4 20 mA Out with Display Loop Loop current current Customer Equipment Loop Power Supply Typ 12 24 VDC Open Collector Pulse In Pulse In Return Pulse Ou...

Page 7: ...that the enclosure sealisnotcrimpedortwisted Donotover tightencornerscrews hand tighten only Faceplate can be rotated 90 W I R I N G D I A G R A M 8 4 20 mA or 0 20 mA Output Customer Equipment Witho...

Page 8: ...4 20 mA and Pulse Output option available only with Reed Switch input W I R I N G D I A G R A M 1 0 4 20 mA or 0 20 mA Output and Pulse Output Customer Equipment With Built in Power Supply Input Reed...

Page 9: ...button for 3 seconds to reset the Batch Total to zero When fluid is flowing through the meter a small propeller icon is highlighted Flowrate Feature Tousethisfeature pressandreleaseDISPLAYuntilFLOWRA...

Page 10: ...ser to tweak the factory calibration by a percent that represents application fluid or plumbing differences 1 To calibrate press and hold the CALIBRATE and DISPLAY buttonsforabout3secondsuntilyouseeFL...

Page 11: ...the transmitter If signal is being received deactivate it by one of the following methods Stop the flow through the line that is being recorded Disconnect the output device on the flowmeter Locking U...

Page 12: ...presently operating temporarily disable its external power supply Be sure to allow at least 30 seconds to elapse with unit unpowered 2 Press and hold the 4 button While holding watch the indicator li...

Page 13: ...signal input connections for intermittent open or short circuits F LOOP OUTPUT 1 Output response time setting too long Select a shorter response time setting STABLE BUT especially for fast input sign...

Page 14: ...14 DIMENSION DRAWINGS...

Page 15: ...Tapping GM001 GM002 and GM003 2 9 120058 01 Bracket 1 10 904005 13 Screw 6 32 x 1 2 in 4 Item No No Part No Description Req d 11 904002 44 Screw 8 32 x 5 16 in 2 12 125066 20 Cable 20 ft 1 125066 3 C...

Page 16: ...linearity plus span any input loop current output 0 C to 70 C loop voltage supply 12 VDC to 24 VDC 0 5 of span plus possible resolution uncertainty Max Conversion Resolution Uncertainty Loop current o...

Page 17: ...gital Pulse Out Open Collector square wave Will switch up to 60 V and up to 200 mA Closed circuit voltage drop typically 0 1 V guaranteed less than 0 5 V at any in range current Configuration 2 Totals...

Page 18: ...Please contact GPI before returning any part It may be possible to diagnose the trouble and find a solution with a telephone call GPI can also inform you of any special require ments you will need to...