![Gow-Mac 75-850-BV Operating Manual Download Page 7](http://html.mh-extra.com/html/gow-mac/75-850-bv/75-850-bv_operating-manual_2248323007.webp)
Model 75-850-BV / 75-852-BV H
2
Separator w/ Manual Bypass Valve
09/20 Rev. 7
7
6. General Specifications for the Temperature Controller
Part
Number
124-261
Control Mode:
- Microprocessor-based
-
Single
input,
dual
output
- Ramp to set point: 0 to 550°
-
Heat
and
cool
auto-tune
- PID with automatic tuning
Sensor Input Type:
Platinum RTD, 100 ohms
Supply Voltage:
85 - 264 VAC, 50/60 Hz (75-850-BV/75-852-BV)
Mechanical Relay Life Span:
100,000 cycles
Sensor Input Type:
Platinum RTD, 100 ohms
Ambient Temperature Range:
Operating: - 18 to 65 °C
Storage:
- 40 to 85 °C
7. Separator Specifications
Operating temperature:
300 °C
Max. Operating Pressure:
250 psig
Operating Carrier Gas Flow Rate: 30-50 mL/min.
Standard Fittings:
1/8" VCR
Weight
(net):
15
lbs.
Dimensions:
15-½”D x 6-½”W x 12"H
8. Waste Disposal
This procedure is for GOW-MAC Mini Purifier Part No. 180-542 (Helium) and
180-542-AR (Argon).
If the purifier is used, then the getter material inside the purifier is already in a non-
reactive state, (i.e. will not react with air).
If the material is still in a reactive state (i.e. the purifier is not completely "used up"),
it still will not sustain combustion in air except at temperatures over 350 °C,
and with a continued access to air.
The operator should place a VCR cap over the inlet/outlet fittings for sufficient
precaution before disposal. It is then acceptable to deposit in an approved landfill,
in accordance with local, state and/or federal regulations. When disposed of in this
manner, the waste poses no known environmental problems.