Model 75-850-BV / 75-852-BV H
2
Separator w/ Manual Bypass Valve
09/20 Rev. 7
3
GOW-MAC Model 75-850-BV or 75-852-BV
Hydrogen Separator
1. Connections
1. Connect AIR CYLINDER to the back of the Hydrogen Separator at AIR fitting.
2. Using the ROTAMETER located on the front panel of the Separator, set the AIR
FLOW to the optimum setting according to the chromatogram located at the back of
this manual. The flow of air can range between 100 - 300 ml/min.
3. Using the 1/16" VCR tubing that is supplied, connect the Separator to the Series 590
DID GC as follows:
THE INLET/OUTLET FITTINGS ARE VCR TYPE AND
WASHERS
MUST
BE USED. CHECK THE CARRIER FLOW
RATE TO BE SURE THAT NOTHING HAS CHANGED.
a. Connect one end of the supplied tubing to "FROM INST" fitting located on the
side of the Model 75-850-BV or 75-852-BV and the other end of the tubing to "TO
SEP" fitting located on the rear of the GC.
b. Connect one end of another piece of the supplied tubing to "TO INST" fitting
located on the side of the Model 75-850-BV or 75-852-BV and the other end of
the tubing to "FROM SEP" fitting located on the rear of the GC.
c. Turn BYPASS VALVE to SERIES/RUN position.