Model 75-800 & 75-800-AR Noble Gas Purifier
Rev. 7 05/19
9
6.2 Responding to a Displayed Message/Error Code
An active message will cause the display to toggle between the normal
settings and the active message in the upper display and [
Attn
] in the lower
display.
Your response will depend on the message and the controller settings. Some
messages, such as Ramping and Tuning, indicate that a process is underway.
If the message was generated by a latched alarm or limit condition, the
message can be cleared when the condition no longer exists. If an alarm has
silencing enabled, it can be silenced.
Push the Advance Key to display [
.9nr
] in the upper display and the message
source (such as [
L.hi
]) in the lower display.
Use the Up and Down keys to scroll through possible responses, such as
Clear
[
CLr
] or Silence [
S.L
]. Then push the Advance or Infinity key to execute the
action.
Er. 1
-
Error Input 1:
The sensor input generated a value lower than the
allowable signal range. Enter a valid input.
100
-
Device Error:
Controller displays internal malfunction message at
Power Up.
7
Set Point Settings
A. To set the set point: turn the unit “on” and then push the Up or Down arrow to
adjust to the desired operating temperature.
8
Settings
8.1 Initial Start-Up
A. The purifier should be purged with an inert gas (minimum purity 99.99%)
before activating the material (He, Ne, Ar, Kr, Xe).
B. Start a flow through the unit for at least 10 minutes. Use a minimum of
10% - 20% of specified flow rate for the unit. Use a continuous purge flow
during activation.
CAUTION
THE PURIFIER MATERIAL SHOULD NEVER BE
HEATED ABOVE 150°C IN AIR.