7.4 Automatic Valve Operation
A. The valves in the SERIES 5900 DID Gas Chromatograph can be operated either
manually by the touchscreen or through the use of the DB9 connectors on the rear of
the GC. The external valve selection on the LCD touchscreen must be selected
to "ON" in order to accept the signals from a remote source or software.
The schematic below shows the pin configuration for the valves in relation to the
DB9 connector. The DB25 connector board can be used to remotely control the GC
and digitally read the output signal. See schematic below for wiring configuration.