GOW-MAC Instrument Co.
10
3.3 Unpacking - Inspection
A. When unpacking the instrument, check it carefully for evidence of shipping
damage or rough handling. Check to ensure that all components ordered have
either been supplied or back-ordered. Notify GOW-MAC of any discrepancies
immediately.
The packing box should be retained for use if the instrument needs to be returned
to the factory for repair or modification. GOW-MAC
does not
supply field service.
ALL
repairs are made at Bethlehem, PA, USA or by an authorized GOW-MAC
representative.
B. Remove all plastic and/or paper shipping caps and restraints before operating.
3.4 Location
A. The SERIES 5900 DID GC should be placed in a location that is secure, vibration-
free, protected from abrupt temperature changes (operating ambient temperature
range is 15 °C - 40 °C), and drafts. Such changes may upset the temperature
stability in the course of an analysis or preparation.
B. Enough adjacent tabletop space should be allowed for the installation of
computers, etc. Allow sufficient space on all sides of the GC for easy access.
C. Make sure that there is adequate space for the installation of the gas cylinders.
Cylinders should be securely fastened to an immoveable structure per CGA and/or
O.S.H.A. regulations.
D. An electrical outlet (AC) should be near the location where the GC will be installed.
If the outlet is not a 3-pin type, make sure that a good ground connection is
available since a good ground is necessary for proper operation. The AC outlet
should be connected to a circuit that is not heavily loaded with other electrical
equipment. Input voltage to the GC should be steady for optimum operating
stability.
If the AC line voltage varies, consideration should be given to the installation of a
stabilizing transformer at the AC outlet. An uninterruptable power supply (UPS)
with surge protection is also recommended to prevent disruption from power loss or
electrical spikes.
Summary of Contents for 5900 Series
Page 2: ...GOW MAC Instrument Co 2 ...
Page 26: ...GOW MAC Instrument Co 26 ...
Page 36: ...GOW MAC Instrument Co 36 ...