GOW-MAC Instrument Co. Series 400 / 402 Gas Chromatograph Operating Manual
May 2017, Rev. 7
42
C.
GC POWER SWITCH "ON"
.
D.
INJECTION PORT TEMPERATURE CONTROL
to 130-135 °C (10-15 °C higher than
the column oven temperature).
E.
COLUMN TEMPERATURE CONTROL
to 120 °C.
F.
DETECTOR TEMPERATURE CONTROL
to 130-135 °C (10-15 °C higher than the
column oven temperature).
WARNING
!
THE DETECTOR MUST BE KEPT APPROXIMATELY 10-15 °C
HIGHER THAN THE COLUMN OVEN TEMPERATURE TO AVOID
CONDENSATION AND CONTAMINATION OF THE TCD.
G. Allow the GC to come to temperature, about 40 minutes.
H. After the GC has warmed up, press the
SELECTOR Button
for
BRIDGE CURRENT
.
I. Turn the
DETECTOR CURRENT SWITCH
to
ON
.
J. Adjust the
CURRENT CONTROL
to 100 mA.
K. Turn the recorder
ON
.
L. Zero the recorder (Chapter 4). If a peak goes off scale, the
ATTENUATOR
should be
turned to a position that brings the peak back on scale.
M. Set
POLARITY SWITCH
for an upscale drift.
8.5 Procedure
A. Temperatures required for this experiment are 130 °C on the
DETECTOR
and 120 °C
on the
COLUMNS.
The
INJECTION PORT TEMPERATURE
should be approximately
10-15 °C higher than the column temperature.
B. Set the
POLARITY SWITCH
to the Carbowax column. Put the
ATTENUATOR
on
"1". When the temperatures have reached the desired settings and the baseline of the
recorder has lined out you are ready to make an injection.
C. Draw a 1 μL SAMPLE into the syringe and then draw about 5 μL of air. With the
recorder running (1"/min. or faster), inject the sample onto the Carbowax column.
Summary of Contents for 400 Series
Page 2: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 2 ...
Page 4: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 4 ...
Page 8: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 8 ...
Page 34: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 34 ...
Page 40: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 40 ...
Page 56: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 56 ...
Page 58: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 58 ...
Page 60: ...GOW MAC Instrument Co 60 ...
Page 62: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 62 ...
Page 64: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 64 ...