A. Gas Flow
!
IMPORTANT
Inert carrier gas
MUST
be fl owing through
both columns A and B as well
as sides A & B
of the TC detector.
THIS IS EXTREMELY IMPORTANT and
NECESSARY to preserve the fi laments!
Failure to set fl ow rates on both sides causes oxidation and destruction of the
fi laments in the presence of oxygen.
1. Make sure that
ALL
switches are in the
OFF
position.
2. Set helium (or other carrier gas) pressure regulator to 40 psig. A minimum of 20 psig
is necessary.
3. Adjust carrier gas to: 30 mL/min. for 1/8" columns
60 mL/min. for 1/4" columns
4. Check for leaks as described in Chapter 3.
5. Disconnect fl owmeter.
6. Allow 5 minutes to purge the system before turning power
ON
.
B. Temperature Controls
NOTE
&
TEMPERATURE CONTROL KNOBS are locking type.
Push locking ring "IN" to turn knob.
1.
ALL
switches should be in the
OFF
position.
2. Turn
DETECTOR CURRENT CONTROL
fully counterclockwise (CCW).
THIS STEP
IS EXTREMELY IMPORTANT!
3. Plug in the GC to the appropriate ac outlet. Switch the instrument ON
(HELIUM MUST
BE FLOWING!).
4. Set
INJECTION TEMPERATURE CONTROL
to desired setting.
5. Set
COLUMN TEMPERATURE CONTROL
to desired setting.
6. Set
DETECTOR TEMPERATURE CONTROL
to desired setting.
Chart 4-1 illustrates the maximum currents for cell temperatures and carrier gases
helium, nitrogen, and argon.
Summary of Contents for 400 Series
Page 2: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 2 ...
Page 4: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 4 ...
Page 8: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 8 ...
Page 34: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 34 ...
Page 40: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 40 ...
Page 56: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 56 ...
Page 58: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 58 ...
Page 60: ...GOW MAC Instrument Co 60 ...
Page 62: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 62 ...
Page 64: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 64 ...