12
GMC-I Messtechnik GmbH
5
How to Display and Store Reports
Note
Storage of the results of the safety and function tests as well as their
entry into reports and statistics is only possible after tests have been
performed.
Exception:
the function switch is in the „MENUE“ position
(older versions: also „FUNCTION-TEST“). In this position
only the result of the last function test can be stored.
You can load the result of the last test into the SI module and store it
there under an Ident Number. The report of the last test can be printed
out several times.
The results of a complete working day (approximately 300 reports) can
be stored in the memory of the SI module. Pressing the
key
several times makes for the same data record to be overwritten again
and again.
The test report automatically contains the test result including the
measured and limit values as well as information on the visual inspec-
tion. Further specific information on device under test, customer and
repair can be integrated in the test report by entries via the keyboard or a
barcode reader (see chapter 7, page 20).
The report is presented on the LCD in several windows.
5.1
How to Display Reports, Enter and Store Texts
Ð
Request the SI menu via the
key
Ð
Select
Protocol
and confirm with
First, the test result including measured and
limit values is shown. The display contains
available data only.
In further windows that can be selected with the
and
keys, you can display information on the
visual inspection as well as enter texts via the key-
board, and barcodes via the barcode reader (see
chapter 7, page 20). A maximum of 24 characters
can be entered into one line.
STORE
MENU