Gossen MetraWatt METRALINE ISO CHECK Operating Instructions Manual Download Page 6

6

GMC-I Messtechnik GmbH

If the 

 symbol is included in the description of the surge pro-

tection device, the manufacturer’s instructions must be observed 
for the respective device type.

The actual measurement is performed as follows:

Select measurement of surge protection devices with the 

R

ISO

/U

SPD

 key, and select the desired sub-function with the 

FUNC

 key (see also description above). Example:

Figure 3.11: Surge Protection Device Measurement Menu, 

DC Function

Connect the test probes to the surge protection device in ac-
cordance with the manufacturer’s instructions.

Figure 3.12: Connection Example

Note

If a voltage of greater than about 10 V is present at the measured 
surge protection device, its value is displayed and the 

START

 key is 

disabled (see section 3.3.1 on page 4). Eliminate the source of 
interference voltage before restarting the measurement.

Trigger measurement by pressing the 

START

 key. Release the 

START

 key as soon as measurement is started. Rising current 

at the surge protection device is displayed by means of a bar 
graph. At the same time, the measuring voltage value appears 
at the display in the 

U

IN

 field. Measurement is ended automat-

ically.

Read voltage measured at the milliampere point.

Note

The device under test may not be disconnected from the 
test instrument as long as the “!” warning symbol is lit up.

After insulation measurement has been completed, any remaining 
residual voltage is displayed as 

U

IN

, which may result from cable 

capacitance. Contact to the device under test must be main-
tained for as long as the capacitive DUT is being discharged via 
the test instrument’s internal resistor. The falling voltage value can 
be observed directly at the 

U

IN

 display. Do not disconnect the 

DUT until voltage 

U

IN

 is less than 25 V!

Comments

Before measuring surge protection devices, disconnect them 
from the installation.

It’s advisable to study the circuit before performing measure-
ment. Surge protection devices are currently equipped with 
integrated interference suppression filters and the like, which 
may influence measurement results.

3.4

Further Device Functions

Select Language, Query Firmware Version

Before selecting a language or querying the firmware version, dis-
connect both test probes from the device under test / measuring 
circuit and switch the test instrument off.

Press and hold the 

R

ISO

/U

SPD

 

key while switching the test in-

strument on.

The firmware version and other service information appears at the 
display, as well as the language selection menu.

Select the desired language with the corresponding key 
(EN = English, CZ = Czech).

Figure 3.15: Language Selection Menu

The instrument is switched back to normal operation after a lan-
guage has been selected.

Measuring Point Illumination with White LED

The LED can be switched on and off by briefly pressing the 

START

 

key.

Note

No voltage may be applied to the test probes.

!

Figure 3.13: Sample Results 

for the 

U

SPD

 

Measurement 
(discharging is 
active: residual 
voltage = 144 V)

Figure 3.14: Further Sample 

Results for the 

U

SPD

 Measure-

ment (DUT is fully 
discharged: resid-
ual voltage = 0 V)

Summary of Contents for METRALINE ISO CHECK

Page 1: ...Operating Instructions METRALINE ISOCHECK Insulation Resistance Tester 3 349 691 03 2 4 13 ...

Page 2: ...ng The device has been stored under unfavorable conditions for a lengthy period of time The device has been handled impermissibly e g has been dropped from a height of 1 meter or more The test instrument does not function as described in these operating instructions if this is the case we recommend resetting the device as described in section 3 5 on page 7 CAUTION Do not touch electrically conduct...

Page 3: ...s are simultaneously recessed and protected In order to charge batteries which have been inserted into the instrument the flexible test probe s connector plug must be removed and the slide must be pushed to the left so that the socket at the right is made accessible and the charger can be plugged into it Figure 2 2 Detail View of the Back Panel with Battery Compartment Lid 2 2 Control Panel 1 Grap...

Page 4: ...nce voltage to appear at the display Do not remove the test probes prematurely during the course of the measurement because this may lead to distorted mea surement results After insulation measurement has been completed any remaining residual voltage is displayed as UIN which may result from cable capacitance Contact to the device under test must be main tained for as long as a capacitive DUT is b...

Page 5: ...y take several tens of seconds Note Whistling or hissing noises may emanate from the test instrument during measurement 3 3 3 Measurements at Varistors surge protection devices For this measuring function the test instrument generates a rising direct voltage within a range of 50 to 1000 V and at the same time measures current flowing through the device under test As soon as current reaches a level...

Page 6: ...hich may result from cable capacitance Contact to the device under test must be main tained for as long as the capacitive DUT is being discharged via the test instrument s internal resistor The falling voltage value can be observed directly at the UIN display Do not disconnect the DUT until voltage UIN is less than 25 V Comments Before measuring surge protection devices disconnect them from the in...

Page 7: ...ast 750 mAh Number of measurements with batteries at 800 mAh approx 1 000 measurements with 500 V test voltage on 500 k Electrical Safety Measuring category with safety cap applied to test probe CAT III 300 V without safety cap applied to test probe CAT II 300 V Pollution degree 2 Protection class II Mechanical Design Display OLED multicolored graphic Protection IP 43 Dimensions Approx 260 x 70 x ...

Page 8: ...d soapy water for cleaning Do not place the test instrument back into service until its surface is completely dry Attention Do not use cleaning agents which contain benzine or al cohol Prevent liquids from penetrating into the test in strument s interior 5 3 Recalibration The measuring tasks performed with your instrument and the stressing it s subjected to influence aging of its components and ma...

Reviews: