Operating Instructions
PROFITEST H+E TECH
Page
13
Errors
Error
Cause
Required Action
After being switched on, the
device switches itself back off
again.
On button not pressed long
enough.
Press and hold the on button
until the device remains on.
The device does not respond
when the on button is pushed.
No batteries are installed or
the batteries are depleted.
Install new batteries or use the
power pack.
The device indicates incorrect
values for PWM voltage.
Battery depleted
Observe the battery display
and install new batteries if
necessary.
Maintenance
Due to the device’s characteristics, no maintenance should be carried out by the user. If repairs
should become necessary, please contact our repair and replacement parts service department
(see address on back cover).
Outside surfaces may only be cleaned with a dry, lint-free cloth.
Danger!
No liquids may be permitted to penetrate into the device’s interior or the plug connectors.
Technical Data
Input voltage
400 V (3-phase)
Frequency
50 Hz
Test consumer power
Max. 2.9 kVA
Electrical Safety
Protection class
I
Nominal voltage
400 V DC
Test voltage
500 V DC
Measuring category
CAT III, 300 V
Pollution degree
2
Ambient Conditions
Operating temperature
-10 to 45 °C
Storage temperature
-25 to 60 °C
Relative humidity
Up to 80% (no condensation)
Mechanical Design
Dimensions (W x L x H)
200 x 240 x 115 mm
Weight
3.65 kg
Protection
IP 21