English
Clamp Multimeter METRACLIP 86
4.2.19 Frequency measurements
4.2.19.1
Characteristics in voltage
Measurement range (1)
5.0 Hz to
599.9 Hz
600 Hz to
5999 Hz
6.00 kHz to
19.99 kHz
Specified measurement
range
1 to 100% of the
measurement
range
0 to 100% of the measurement
range
Uncertainties
± (0.4% R + 1 pt)
Resolution
0.1Hz
1Hz
10Hz
4.2.19.2
Characteristics in current
Measurement range (1)
5.0 Hz to 2999 Hz
Specified measurement range
1 to 100% of the measurement range
Uncertainties
± (0.4% R + 1 pt)
Resolution
0.1Hz
Note (1) in MAX/MIN mode, the operating range is limited to 1kHz.
- If the level of the signal is too low (<10% of the range, or U<10V or
I<10A) or if the frequency is less than 5Hz, the device cannot determine
the frequency and displays dashes "----".
Specific characteristics in MAX/MIN mode
(from 10Hz to 1kHz):
•
Uncertainties: add 1% R to the values of the table above.
•
Capture time of the extrema: approximately 100ms.
4.2.20 Characteristics in THDr
Measurement range
0.0 – 100%
Specified measurement range
0 to 100% of the measurement range
Uncertainties
± (5% R ±2 pts) in voltage
± (5% R ±5 pts) in current
Resolution
0.1%
4.2.21 Characteristics in THDf
Measurement range
0.0 – 1000%
Specified measurement range
0 to 100% of the measurement range
Uncertainties
± (5% R ±2 pts) in voltage
± (5% R ±5 pts) in current
Resolution
0.1%
45