WG600S Industrial Basic Variable-Speed Peristaltic Pump 

3

 

Components and Connectors 

8

.

8

.

8

.

8

Digital LED

SPEED ADJUST Dial

PRIME/DISPENSE Button

DIRECTION Switch

Pump Head Holder

Coupler

Power Switch

Cooling Fan

Power Connector

DB15 External 

Control Interface

 

 

Figure 1. Components and Connectors 

3

 

Summary of Contents for WG600S

Page 1: ...GOLANDER PUMP Operation Manual For Industrial Variable Speed Peristaltic Pump WG600S 8 8 8 8 info golanderpump com http golanderpump com 1 678 587 8806...

Page 2: ...tion Switch 7 5 External Control Interface 7 6 Operation Instructions 8 6 1 Before Operation 8 6 2 Power Connection 8 6 3 Internal Control Mode 8 6 4 External Control Mode 9 6 5 Time Dispense Mode 11...

Page 3: ...from pump before connecting or disconnecting the external control device or communication interface Warning Pump is provided with a grounded plug it must be well grounded at all times Warning This pr...

Page 4: ...materials are available for food and pharmaceutical use Four digit LED displays speed Reversible direction start stop control and adjustable speed Time dispense mode to dispense fluid for the set time...

Page 5: ...ic Pump 3 Components and Connectors 8 8 8 8 Digital LED SPEED ADJUST Dial PRIME DISPENSE Button DIRECTION Switch Pump Head Holder Coupler Power Switch Cooling Fan Power Connector DB15 External Control...

Page 6: ...Buttons 8 8 8 8 RPM Display DIRECTION Switch PRIME DISPENSE Button SPEED ADJUST Dial Figure 2 Display and Buttons 4 1 Digital LED Display Digital LED display shows the current speed and working mode...

Page 7: ...Pump E100 Figure 5 External Control Mode 100 rpm H215 Figure 6 Logic Level Control Mode 215 rpm l215 Figure 7 Communication Mode 215 rpm C 4 9 Figure 8 When flashing the pump is working on Time Dispen...

Page 8: ...te the dial in counterclockwise direction to decrease the setting Rotate 1 position to decrease 1 to the last digit Rotate continuously to decrease the setting fast Press the dial to switch the contro...

Page 9: ...ton to dispense fluid 4 4 Direction Switch When on left position the drive runs in counterclockwise direction When on middle position the drive stops When on right position the drive runs in clockwise...

Page 10: ...e package If there is problem please contact the manufacturer or distributor 2 Read through the instruction 3 There should be more than 200 mm space for the back of the pump when it is running 6 2 Pow...

Page 11: ...on the drive will run at maximum allowed speed 7 Change DIRECTION switch to the middle position to stop the drive anytime when an accident occurs 215 fUll PRIME DISPENSE Figure 12 Full Speed 6 4 Exter...

Page 12: ...and connect it to the DB15 port on the rear of the pump 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 ADC_W AGND 0 5V 0 10V 4 20mA 12VDC VCC_W COM RS_W CW_W Figure 14 DB15 Wiring with External 12VDC Power Sourc...

Page 13: ...lose the CW_W switch then the drive will run in counterclockwise direction E 0 E125 Drive Stopped Drive Running Figure 16 External Control Mode Note The external DC power source can be 5V 12V or 24V I...

Page 14: ...up Mode use the SPEED ADJUST dial to set the duration time the time range is 0 1 999 seconds 6 Press the PRIME DISPENSE key again to exit the Time Dispense Setup Mode To Dispense Fluid 1 Press SPEED A...

Page 15: ...ADJUST dial to change the mode to Logic Level Control Mode 4 Rotate the SPEED ADJUST dial to adjust the speed to desired value 5 Change DIRECTION switch position to desired rotating direction 6 When...

Page 16: ...10 11 12 13 14 15 12VDC VCC_W COM B A Figure 21 RS485 MODBUS Wiring 2 Turn on the power switch The LED display will be on 3 Press SPEED ADJUST dial to change the mode to Communication Mode 4 Change DI...

Page 17: ...ress SPEED ADJUST dial to change the mode to Time Dispense Mode The RS_W would act like a momentary switch When the switch is pushed then released the pump will dispense one dose 4 Press SPEED ADJUST...

Page 18: ...ow fuse 3 Check the internal power cord connection inside the pump 2 Hardware Motor does not work 1 Check the DIRECTION switch if it is on the middle position 2 Check the indicator of the driver board...

Page 19: ...connector 2 Check if the external control power voltage is provided 3 Check the connections of the communication board 4 Check the DIRECTION switch if it is on the middle position 8 Hardware Noisy wh...

Page 20: ...9 for special pumps Version 0 9 Rotation Speed 05 50 rpm 08 80 rpm 10 150 rpm 30 350 rpm 60 600 rpm Flow Rate Q Microflow T Laboratory G Industrial P Batch transfer IP rating Default IP31 Pump type D...

Page 21: ...ional Communication interface RS485 MODBUS Operating condition Temperature 0 40 C Relative humidity 80 IP grade IP31 Display Four digit LED Dimensions LxWxH 350x235x230 mm 13 78x9 25x9 06 inch Weight...