WG600S Industrial Basic Variable-Speed Peristaltic Pump
3) Press
SPEED ADJUST
dial to change the mode to
External Control Mode.
4) Close the external RS_W switch, and turn on the external
analog signal power source. The speed will change according to the
intensity of the input signal. Open the RS_W to stop the drive.
5) Open CW_W switch, then the drive will run in clockwise direction; close
the CW_W switch, then the drive will run in counterclockwise direction.
E
0
E125
Drive Stopped
Drive Running
Figure 16. External Control Mode
Note
: The external DC power source can be 5V, 12V or 24V. If it is 24V, 1.5K
resistor is needed to protect internal circuit.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
ADC_W
AGND
0-5V/0-10V
/4-20mA
24VDC
VCC_W
COM
RS_W
CW_W
1.5K
1.5K
Figure 17. DB15 Wiring with External 24VDC Power Source
6.5
Time Dispense Mode (F)
Pump will dispense fluid by setting the duration time for each dose. The drive
will stop automatically when finished dispensing.
E215
11