BT-F NEW Intelligent Dispensing Peristaltic Pump
25
3) Press prime key to fill tubing with liquid.
4) When pump is not running, press
, or enter system settings to
choose calibration wizard icon.
5) On the calibration wizard screen, system shows the calibration of
current selected tubing, flow rate and fluid volume.
Calibrate
Analyze and Calculate,No data.
Please go back to re- enter the
data or return to the main menu.
3/3
Figure 39 Flow Rate Calibration
The flow rate 9.000 mL/min is the desired flow rate and 6.000 mL is the
fluid volume need to test. The values or the units can change directly
when press the button. Press
button to enter the calibration
window, or press
button to exit the wizard.
Note
: The fluid volume should not be less than the suggested value.
6) Test window shown as below.
Calibrate
Press START/STOP key to test,
then input the data.
2/3
Test1:
Test3:
Test2:
mL
0.000
0.000
0.000
mL
mL
Calibrate
Press START/STOP key to test,
then input the data.
2/3
Test1:
Test3:
Test2:
mL
0.000
0.000
0.000
mL
mL
Figure 40 Calibration
Press the start/stop key, the peristaltic pump will start to transfer fluid.
Wait for the pump to finish testing, and then measure the delivered