BlueSense - Multipoint calibration
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 52 / 55
www.go-sys.de
3. Input of the calibration values
Immerse the sensor into the calibration medium. Enter the turbidity value of the calibration medium.
Aborts the calibration and switches back to the turbidity menu.
Deletes the last entered character.
Saves the input and switches to the next menu.
Without a setting the value is set to 0. Not recommended!
Aborts the calibration and switches back to the turbidity menu.
Saves the in calibration and switches back to the turbidity sensor menu.
The calibration is saved and completed.
1/6
actual number of the measurement point / number of measurement points
0.0
actual measurement value and therefore reference value, here 0.0
FNU
unit of the measurement, here FNU for turbidity
4. Finalizing the calibration
List menu:
List of Reference values
List of Input values