CORRECTIVE AND PREVENTIVE MAINTENANCE AND
PRESERVATION
Clearing the suction system
Using Vacuum Pump (BioVac II or IV):
Using Vacuum Pump (BioVac Sec):
Gnatus suggests performing a daily suction of the clearance and
disinfectant solution, avoiding the risk of cross contamination and
increasing equipment service life. To perform the disinfection of your
equipment we recommend the use of the “Sugclean” (MS Reg. No.:
31.080.003-2) product.
•
Indication:
It is indicated for clearance of sucker and hose
suction system. It is important to perform the suction solution in all
suction terminals, which it is also important to be open. Then, remove
suckers from hose for asepsis (Fig.A).
•
Preparing the Solution:
Add “Sugclean” 30mL in 1 liter of
water. Aspirate the solution with maximum power of the suckers, and
also put the liquid in the water unit bowl.
In the first use of “Sugclean” product, we suggest adding
60mL of concentrated product in 1 liter of water during the first
5 days in order to remove accumulated residues.
•
Composition:
• Active Drug: Phosphoric Acid 13.6%
• Excipients: Isopropyl Alcohol, Acidulant, Dye and Thickener.
Warning:
do not use foaming product.
NOTE:
The registration at the Ministry of Health of the
“Sugclean” is executed separately from the product described in
this manual, as the “Sugclean” is not manufactured by Gnatus.
CAUTION:
For internal cleaning of Suction pipelines of the Pump Vacuum BioVac Sec, the
use of "Sugclean" product or any similar product is not allowed; use only the recommended
mixture below:
• After each patient leaves, perform the suction of 250 ml of clean water in each totally
open suctor;
• At the end of the working Day, perform the suction of the 250 ml mixture of bleach
mixed with 250 ml of clean water (proportionally divided in each used suctor); Then, remove
suckers from hose for asepsis (Fig.A).
WARNING:
• Never use foamy products in the suction (deep cleaner, detergents, floaters, etc), this
procedure may damage the internal parts of the dry vacuum pump’s engine;
• Never use the bleach solution for external cleaning of any equipment, because this
mixture is highly corrosive and may damage metal parts.
Fig.A