20
• For your safety the Optilight LD Max is equipped with an automatic bi-voltage power supply of
90V~-240V~ - 50/60Hz.
• Plug in the power supply (01) to the electrical outlet.
• Select application time, press time selection button (10), which values are: 10s (standard mode),
20s, 60s, 80s and 90s.
• To initiate a polymerization cycle, press the timer trigger (09), which generates a short beep every
10 seconds and a 4 beeps at the end of the cycle.
• To interrupt a polymerization cycle just activate the timer trigger again (09).
IMPORTANT:
• Keep the light conductor tip (06) at least 2mm away from the restoration.
• Keep the light conductor (06) always protected by an expendable PVC film, which must be changed
for every patient. This procedure protects the light conductor from scratches and other residues.
• Use the polymerization time recommended by the compound resin manufacturer and always
perform restorations in incremental layers with a maximum thickness of 2mm.
Instructions for equipment operation
WARNING
• Never aim the blue light beam towards the eyes
• Use the eyesight protection (05)
In order to protect the eyes, the eyesight protection (05) filters only the blue light used for the resins
polimerization, and it allows ambient light to pass through.
OPERATION