GMI D1010S-054 Instruction Manual Download Page 7

 

D1010-054 / 056 / 057

- mV to mA Converter 

G.M. International ISM0098-8 

Configuration 

Span 

Screwdriver 

for Trimmers 

1 x 0.5 mm 

Calibration Trimmers Channel 

Zero 

Turn the trimmer clockwise to 
Increase Output value or turn 

the trimmer counterclockwise 

if you want to decrease Output 

value. 

Side A Panel View 

 

Calibration instruction and Troubleshooting 

In case a calibration check or readjustment is necessary the following equipment is required: 

PS

Supply

=

_

Out 1

+

CAL

In 1

DMM mA

mV

MODELS D1010S-054, D1010S-056, D1010S-057

=

=

=

=

=

15

16

3 +

4 -

1

2



Connect the unit as shown below: 



Set power supply (PS) at 24.0 V at terminals 3 (+) and 4 (-), the calibrator (CAL) in the function “mV” 100 mV range at terminals 15 (-) and 16 (+), the multimeter (DMM) for the reading 
in DC supply 20 mA range at terminals 1 (+) and 2 (-). 



Verify that green led on the front panel is ON. 

If the power ON Led is off, check supply voltage, polarity and wiring. 

Adjustable Power Supply Range 0-35 V Resolution 0.1 V, 
Accuracy 1% Output Capability 200 mA 
Current Meter range 20 mA, Resolution 1 µA Accuracy 0.01% 
or Voltage Meter Range 10 V, Resolution 1 mV Accuracy 0.01% 

Current Calibrator range 100 mV, Resolution 1 µV Accuracy 0.01 % 

Agilent  

E3611A 

Agilent 

Yokogawa 

7651 

Equipment Required 

Instrument 

PS 

DMM 

CAL 

Identification 

Summary of Contents for D1010S-054

Page 1: ...D1010 054 056 057 mV to mA Converter ISM0098 8 D1010S 054 D1010S 056 D1010S 057 INSTRUCTION MANUAL mV to mA Converter DIN Rail Models D1010S 054 D1010S 056 D1010S 057...

Page 2: ...oad 23 1 C ambient temperature Calibration accuracy 0 1 of full scale Linearity error 0 1 of full scale Supply voltage influence 0 02 of full scale for a min to max supply change Load influence 0 02 o...

Page 3: ...gh Accuracy Three port isolation Input Output Supply EMC Compatibility to EN61000 6 2 EN61000 6 4 ATEX IECEx Certifications High Reliability SMD components Simplified installation using standard DIN R...

Page 4: ...Co Ca 100 F Co Ca 1000 F Co Ca 1000 F IIB C IIA D IIC A B Lo La 11 3 mH Lo La 45 3 mH Lo La 90 7 mH IIB C IIA D D1010 054 D1010 056 D1010 057 Associated Apparatus Parameters D1010 054 D1010 056 D1010...

Page 5: ...mA Converter G M International ISM0098 8 Function Diagram HAZARDOUS AREA ZONE 0 ZONE 20 GROUP IIC SAFE AREA ZONE 2 GROUP IIC T4 MODELS D1010S 054 D1010S 056 D1010S 057 15 16 3 4 1 2 RL mA Supply 24 V...

Page 6: ...10 05 must be cleaned only with a damp or antistatic cloth Any penetration of cleaning liquid must be avoided to prevent damage to the unit Any unauthorized card modification must be avoided According...

Page 7: ...3 4 1 2 Connect the unit as shown below Set power supply PS at 24 0 V at terminals 3 and 4 the calibrator CAL in the function mV 100 mV range at terminals 15 and 16 the multimeter DMM for the reading...

Page 8: ...35 000 mV and regulate the current read by multimeter DMM by turning the span trimmer of the unit see Configuration section until reading 20 000 mA 5 A value Set the calibrator CAL at 5 000 mV and reg...

Reviews: