45
EN
6
Technical
Specifications
Emission:
Measurement spot size: 4.5/8 mm
Measurement range: 0…300 cd/m2
Video frequency: 50 – 250 Hz
Emission spectra:
spectral radiometric emission measurement
Colorimetry: XYZ, xyY
Standard observers: 2
°
, 10
°
Brightness base: absolute, relative
Measurement time: ca. 1.5 s
Short-term repeatability: x, y:
±
0.001
(typically, measured on white 80 cd/m2)
Transmission (with SpectroScan T):
Measurement geometry:
diffuse/specular (d/0), specular/specular
(0/180) to DIN 5033
Apertures: 1 mm, 2 mm, 3 mm
Illuminant: illumination A to ISO 5/3
Spectra: transmission, density
Colorimetry:
CIE-XYZ, CIE-xyY, CIE-L*a*b, CIE-
L*C*h (a*b), CIE-L*u*v, CIE-L*C*h*(u*v*),
LABmg, LCHmg, HunterLab
Illumination types:
D50, D65, A, C, D30… D300, F1…F12
Standard observers: 2
°
, 10
°
Density standards:
DIN 16536, DIN 16536 NB, ANSI STATUS A,
ANSI STATUS T
Measurement time: 5 s
Short-term repeatability:
0.5
∆
E*CIELab (D65, 2
°
)
Mean value of 10 measurements every
15 s on 3 mm aperture, diffuse
Measurement range:
Standard deviation of 10 measurements
every10 s with 3 mm aperture, diffuse/
specular:
Density (ANSI A):
Cyan
0.0 D – 2.5 D : 0.03 D
Magenta 0.0 D – 2.5 D : 0.03 D
Yellow
0.0 D – 2.5 D : 0.03 D
Data interface specification:
RS 232C serial interface with baud rates from
110 to 57,600
Data interface connector: Mini-Din connector
Power supply: 15 V DC 0.8A
Power supply adapter specifications:
85 VAC – 270 VAC, 47 – 63 Hz