40
EN
5
SpectroScan T
operation
5.4.3 Supplementary functions
5.4.3.1 Reset
The SpectroScan T software may be reset by pressing the and
keys. Resetting reinitialises the measuring table.
5.4.3.2 Demo Mode
Keeping the and depressed when switching the measuring table on,
simulates a test chart measurement. The measuring table crosses 20 fields trans-
versely, and 10 fields longitudinally. The procedure may be interrupted by switching
the table off.
5.4.3.3 Replacing fuses
First, disconnect the device from the main power supply by removing the plug. In the
storage compartment (45) of the SpectroScan T there are two spare fuses. To replace
these fuses, open the fuse box (53) on the back of the device. The measuring table
may only be switched on again when the fuse box is closed.
Enter
Online
- Before opening the fuse box, always disconnect the device from the main
power supply. Fuses must be replaced with code 13.52.05 type fuses
supplied by GretagMacbeth .
- Resetting must be performed in cases of connection problems between
user software and the SpectroScan T.
- Resetting of the measuring table is independent of the Spectrolino. If a
communication anomaly occurs, the measuring device must be reset
before the reset of the measuring table (see chap. 3.4.3).