Electronic monitoring equipment (such as ECG monitors and alarms) may not
operate properly when electrostimulation is in progress.
-Stimulation delivered by this device may be sufficient to cause electrocution.
Do not apply the current in trans-thoracic modality because it may cause cardiac arrhyth-
mia. Do not stimulate the pectoral and dorsal muscles simultaneously.
-Do not use the electrostimulator near short wave therapeutic devices.
APPROPRIATE USES
Read the operating manual carefully in order to avoid incorrect installation, use or maintenan-
ce that can damage the apparatus, the operator or the user. The Globus Electrostimulator is
designed to be used in the following settings:
-Home
-Medical clinics
-Physiotherapy clinics
-Rehabilitation clinics
and for the following purposes:
-Pain treatment in general
-Any type of sequential electrostimulation
-Aesthetic care
-Sports training
-The Electrostimulator may be used with or without the supervision of physicians or
paramedic personnel.
EMS and TENS:
Channels available
Channels 1, 2, 3, 4
Constant current
Yes
Intensity
0-120 mA with 1000 Ohm load
Wave form
Rectangular, biphasic, symmetrical, compensated
Working frequency
0.3-150 Hz
Recovery frequency
0.3-150 Hz
Impulse amplitude
50-450 m s
Working time
1-30 s
Recovery time
0-1 min
Frequency modulation range
Continuous variation from 1 to 150 Hz
Minimum modulation time, 3 s
Amplitude modulation range
Continuous variation from 50 to 450 ms
Minimum frequency, 1 Hz
Maximum frequency 150 Hz
Microcurrents MCR :
Channels available
Channels 1, 3
Constant current
Yes
Minimum frequency
0.2 pps
Maximum frequency
200 pps
Minimum intensity
0 mA, 1000 Ohm; 10 mA steps
Maximum intensity
800 mA
Amplitude
1-250 s