VI.

 

Programming 

 

a.

 

Programming Overview 

The programming ring can be rotated from the Neutral center position to 
Position 1 and Position 2. The following actions are possible: 
 
A – Display of parameters with Position 1 (simultaneous display of the 
set parameters) – 

Turn the programming ring left to Position 1 to begin 

cycling through these programming parameters: Switching points S1 and 
S2, Hysteresis direction of S1 and S2, Hysteresis Hyst 1 and Hyst 2, Code 
(allows editing of additional parameters), Filter, Units, Output, 4 mA Value 
and 20 mA value. See following pages for detailed programming 
instructions. 
 

B – Editing with Position 2 

Turn the programming ring to the right to 

Position 2 and a flashing cursor appears showing the position to be 
changed. With repeated turning to Position 2, the values are increased. By 
turning to Position 1, you obtain the next position. Each position can be 
edited in this way. If there is no action within 5 seconds, the device 
returns to the normal display section without the change being 
accepted, and you will have to cycle through the program again. 
 

C – Saving the change with Position 1 

Turning one time toward Position 

1 after quitting the last value signifies acceptance of the change. 

 

b.

 

Programming Protection 
The programming ring can be pulled off, inverted, and replaced. This will 
prevent further programming resulting from turning the ring in either 
direction. (See Protecting your Programming Parameters section.) 

 

 

- 8 - 

 

Summary of Contents for WL750

Page 1: ...er 800 876 1172 globalw com Global Water Instrumentation Inc 11390 Amalgam Way Gold River CA 95670 T 800 876 1172 Int l 916 638 3429 F 916 638 3270 Ultrasonic Level Transmitter WL750 4 18 07 P N 01 73...

Page 2: ...are confident that you will find the Ultrasonic Level Transmitter to be a valuable asset for your application Should you require assistance our technical staff will be happy to help Table of Contents...

Page 3: ...Level Transmitter was carefully inspected and certified by our Quality Assurance Team before shipping If any damage has occurred during shipping please notify Global Water Instrumentation Inc and file...

Page 4: ...everse polarity proof Hysterisis Adjustable position depends on min or max setting Display Graphical LCD display temperature range 20 C to 70 C 4 F to 158 F 32x16 pixels back lit LED signaling lamp Co...

Page 5: ...66 ft to 4 3 ft WL750 8 0 98 ft to 7 5 ft Mechanical Wetted Materials PET 30 GV epoxy resin POM Housing Materials Stainless steel tempered glass POM ring Housing Flange mount with center hole 35 5mm a...

Page 6: ...be set separately in value and direction minimum maximum switching value Upward and downward crossings of switching points and error messages are shown in the display with a flashing red LED that is e...

Page 7: ...o be monitored If the vessel has a sloped top consider using a self aligning bulkhead fitting or a flange fitting with a riser that will allow the sensor to be positioned relative to the material s su...

Page 8: ...iting with Position 2 Turn the programming ring to the right to Position 2 and a flashing cursor appears showing the position to be changed With repeated turning to Position 2 the values are increased...

Page 9: ...ve status is also indicated by a flashing LED By rotating the program ring to Position 1 P1 toward the left and then back to Neutral the program steps to the S1 setting The level for activation of the...

Page 10: ...Global Water 800 876 1172 globalw com 2 Programming Positions The P1 to Neutral movement steps the program through positions in the following sequence Figure 5 10...

Page 11: ...ck to Neutral increases the current decimal place to the next highest value c Turning the ring to the left and back to Neutral moves the cursor on to the next decimal position Repeat step 2 to edit th...

Page 12: ...osition Figure 7 Figure 8 Figure 9 Unlocked Position Locked Position 6 Factory Settings a Enter code 989 to return to factory settings Note During the programming process if there is no new action tak...

Page 13: ...Global Water 800 876 1172 globalw com Figure 10 13...

Page 14: ...Global Water 800 876 1172 globalw com 14 Figure 11 Figure 12...

Page 15: ...e event that the equipment needs to be returned to the factory for any reason please call to obtain an RMA Return Material authorization Do not return items without an RMA displayed on the outside of...

Page 16: ...option I replacing or II repairing any products determined to be defective In no case shall Global Water s liability exceed the products original purchase price This warranty does not apply to any eq...

Reviews: