PART THREE - OPERATION
SECTION 7 - DIAGNOSTICS MENU
Rev. 7-1201
Model P63 pH Analyzer
93
7.4 CALIBRATION
RECORD
(last calibration only)
7.5 SENSOR STATISTICS
(1st and last 3
calibrations)
The analyzer creates a record containing information about
the last calibration. This includes the calibration date and
time, the calibration mode (single or two-point), the sensor’s
zero and slope, the buffer values used, the sensor’s mV
output, and the temperature during calibration.
After selecting “Calibration Record” from the DIAGNOSTICS
menu, a screen similar to this example screen appears:
The analyzer creates sensor statistics records for the first
calibration and the last three (most recent) calibrations. The
data contained in these records is similar to the data in the
calibration record.
After selecting “Sensor Statistics” from the DIAGNOSTICS
menu, a screen similar to this example screen appears:
Items in the “FIRST CALIBRATION” record include the cali-
bration date and time, the sensor’s zero and slope, and
impedance of the measuring and reference electrode. The
other calibration records (LAST CALIBRATION, 2 CALI-
BRATIONS AGO, and 3 CALIBRATIONS AGO) contain the
same information, but it is displayed as the difference be-
tween that respective calibration and the first calibration.
For example, suppose the FIRST CALIBRATION slope was
59.6 mV/pH and the slope for the last calibration was 57.3
mV/pH. The displayed slope for the LAST CALIBRATION
would then be “-2.3 mV/pH” (the negative difference be-
tween the first and last calibrations).