PART TWO - INSTALLATION
SECTION 1 - UNPACKING
Rev. 7-1201
Model P63 pH Analyzer
19
P A R T T W O - I N S T A L L A T I O N
S
ECTION
1
After unpacking, it is recommended to save the shipping
carton and packing materials in case the instrument must be
stored or re-shipped. Inspect the equipment and packing
materials for signs of shipping damage. If there is any evi-
dence of damage, notify the transit carrier immediately.
S
ECTION
2
2.1 Location
1. It is recommended to locate the analyzer as close as
possible to the installed sensor:
•
GLI Differential Technique Sensor:
The maximum
allowable distance between this type of sensor and
the analyzer is 3000 feet (914 m). However, dis-
tances greater than 500 feet (152 m) may degrade
the electrode impedance diagnostics of the ana-
lyzer.
•
Conventional Combination Electrode:
The maxi-
mum allowable distance between this type of
electrode and the analyzer is 100 feet (30 m). A
preamp may be used to extend this distance to 3000
feet (914 m), but the preamp must be within 100 feet
(30 m) of the electrode. (
Important:
A preamp will
eliminate the electrode impedance diagnostics ca-
pability of the analyzer.)
2. Mount the analyzer in a location that is:
➥
Clean and dry where there is little or no vibration.
➥
Protected from corrosive fluids.
➥
Within ambient temperature limits (-22 to +140
°
F;
-30 to +60
°
C).
CAUTION:
EXPOSING THE ANALYZER TO DIRECT
SUNLIGHT MAY INCREASE THE OPERATING
TEMPERATURE ABOVE ITS SPECIFIED
LIMIT.
UNPACKING
MECHANICAL REQUIREMENTS