PART THREE - OPERATION
SECTION 5 - TEST/MAINTENANCE
Rev. 6-1100
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
71
5.10 RESET CONFIGURE
Values to Factory
Defaults
5.11 RESET CALIBRATE
Values to Factory
Defaults
You can simultaneously reset stored configuration settings
to factory-set defaults (see Table B).
This excludes cali-
bration settings.
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to display the “RESET CONFIGURE:
ARE YOU SURE?” screen, asking if you really intend to
perform this extreme action. (If you want to abort this
action, press
ESC key
now.)
3.
Press
ENTER key
to reset all stored configuration set-
tings -- not calibration settings -- to factory defaults
(“RESET CONFIGURE: DONE” screen appears, ac-
knowledging reset has occurred).
4.
To return to the previous level of the TEST/MAINT
menu branch, press
ESC key
or
ENTER key
.
You can conveniently reset stored calibration settings to
factory-set defaults.
This excludes all other configuration
settings.
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to display the “RESET CALIBRATE:
ARE YOU SURE?” screen, asking if you really intend to
perform this extreme action. (If you want to abort this
action, press
ESC key
now.)
3.
Press
ENTER key
to reset all stored calibration settings
-- not configuration settings -- to factory defaults
(“RESET CALIBRATE: DONE” screen appears, ac-
knowledging reset has occurred).
4.
To return to the previous level of the TEST/MAINT
menu branch, press
ESC key
or
ENTER key
.