EnSet
©
Plus / Original / Compact
Version 03 - 15.10.2018
34/40
7.3
Handpiece
EnSet
©
handpiece incorporates a cylinder/ piston system
(24)
. In order to keep
maximum power, replacing this mechanism each 18-24 months is recommended,
depending on amount of activity.
Lubricate the handpiece once a year with vaseline.
No.
Action
(A)
Unscrew the handpiece
(24)
in the middle to access the striker head shaft.
(B)
Remove striker head
(15)
on shaft by loosening the setscrew
(16)
.
(C)
Remove the handpiece cap
(26)
to the cylinder / piston system
(25)
.
(D)
Replace the cylinder / piston system
(25)
.
(E)
Reverse above procedure steps to rebuild.