background image

8. Troubleshooting and Maintenance Services

Please call the after-sales telephone and contact after-sales service personnel of 

MELING BIOLOGY & MEDICAL for replacement. Please do not replace the parts by 

yourself.

The following conditions are not faults

When the compressor starts and stops, the equipment parts will make a slight impact sound;

When overheated items are put in after the door is opened, high temperature alarm and high humidity alarm will

be triggered in the control system (if this function is available, please refer to the alarm display table);

Solution: 

Put the items into the equipment after they are naturally cooled. Put the items in small quantities in 

batches, and do not put too much at a time. After the system runs stably, the high temperature alarm and high humidity 
alarm will be released automatically. 

Slight sound of running water caused by refrigerant flowing in the pipeline.

¿

  The equipment can only be repaired, maintained or improved by the engineers certified by MELING BIOLOGY & 
MEDICAL, so as to ensure the normal operation of the equipment and the compliance with corresponding safety 
standards.

¿

Please clean and disinfect the equipment before notifying the maintenance engineer; During the warranty period of 
the equipment, the Company will not undertake the warranty obligation if the fault or damage is caused by improper 
use of the user.

¿

Ambient temperature for storage: -40°C-+55°C, relative humidity: 10% ~ 90%.

Any product may fail. Please observe the operation of equipment in time during use. If there is any abnormality, please 

check and handle it according to the following table first. If the abnormality can't be changed, please inform our service 

center in time, and we will serve you wholeheartedly to avoid losses.

Fan motor replacement time: about 6 years; Service life of the product: 10 years.

9. Specifications

Notes:

20

21

After the equipment is used for a period of time, the door seal becomes hard and deformed.

Maintenance method: Blow the deformed part of the hot seal with a blower to soften it, and 

then close and compress it after the door seal becomes soft.

Is the power outlet charged?

Is the power plug plugged in or loose? 

Is the power fuse disconnected? 

Is the supply voltage too low or too high?

Is the temperature set correctly? 

Is the temperature inside the cabinet too low?

Are the doors not tightly closed or open too many times? 

Are there too many items put in at one time? 

Is the ambient temperature too high?

Is the cabinet placed on a flat ground? 

Does the cabinet contact the wall?

In rainy and humid seasons, door condensation is normal, and it shall be wiped off 

with a dry cloth.

Equipment needs to be cleaned.

Items with heavy smell are not packed.

Is the temperature set correctly? 

Is the temperature inside the cabinet too low?

Equipment does not 

work

Compressor is not 

running

The temperature 

does not reach the 

set value

High noise

Undesirable odor

The door is not 

closed properly, and 

the cool air leaks

Condensation on 

cabinet surface

The alarm lamp 

flashes and the 

buzzer is triggered

Lighting damage 

(refrigeration 

equipment)

Problems

Causes and solutions

*  The foaming material of this product is cyclopentane.

Model

Ambient 

temperature

Climate 

type

Refrigerant 

and loading 

amount

Rated 

voltage 

(v~)

Rated 

frequency

(Hz)

Temp 

Range

(°C)

Effective 

volume

(L)

Input 

power

(W)

Weight

(Kg)

Exterior 

dimensions 

(DxWxH) (mm)

YC-55L

16 ~32°C

N

R600a/14g 220~230V

50

2~8 °C

55

121

35

560X540X632

YC-55EL

16 ~32°C

N

R600a/14g 220~230V

50

2~8 °C

55

125

32

560X540X632

YC-75L

16 ~32°C

N

R600a/18g 220~230V

50

2~8 °C

75

147

41

560X540X764

YC-75EL

16 ~32°C

N

R600a/18g 220~230V

50

2~8 °C

75

135

35

560X540X764

YC-130L

16 ~32°C

N

R134a/110g 220~230V

50

2~8 °C

130

155

51

625X650X810

YC-315L

16 ~32°C

N

R600a/22g 220~230V

50

2~8 °C

315

285

87

673X650X1762

YC-395L

16 ~32°C

N

R600a/25g 220~230V

50

2~8 °C

395

290

95

673X650X1992

YC-395EL 16 ~32°C

N

R600a/25g 220~230V

50

2~8 °C

395

250

89

673X650X1992

YC-400L

16 ~32°C

N

R290/65g 220~230V

50

2~8 °C

400

500

116

645X700X2016

YC-525L

16 ~32°C

N

R290/75g 220~230V

50

2~8 °C

525

548

148

720X810X1961

YC-725L

16 ~32°C

N

R290/85g 220~230V

50

2~8 °C

725

790

171

758X1093X1972

YC-1015L 16 ~32°C

N

R290/140g 220~230V

50

2~8 °C

1015

796

223 900X 1180X 1990

YC-1015EL 16 ~32°C

N

R290/140g 220~230V

50

2~8 °C

1015

667

206 900X 1180X 1990

YC-1505L 16 ~32°C

N

R507/500g 220~230V

50

2~8 °C

1505

1400

322

830X1795X1990

Summary of Contents for GL-RG-12M

Page 1: ...GL RG 4M GL RG 12M GL RG 16M GL RG 23M OPERATION MANUAL Applicable product model www aegisfridge com ...

Page 2: ...roduct composition and overview 6 Operating Instructions 6 1 Function Introduction 6 2 Optional function 7 Maintenance and Service 7 1 Equipment Maintenance 7 2 Equipment Discontinuation 7 3 Maintenance Replacement and Recovery of Rechargeable Batteries 8 Troubleshooting and Maintenance Services 9 Specifications 10 Packing List 11 Electrical Schematic Diagram 1 2 4 4 4 4 5 5 6 7 10 10 19 19 20 19 ...

Page 3: ...er products For your safe and convenient use and reasonable maintenance of the equipment please read the Operation Instructions carefully before use and keep it properly for reference The equipment operator can copy some chapters of this operation manual but only for internal use for example to instruct the user how to deal with emergencies These chapters are clearly marked in the catalog of the m...

Page 4: ...utlet tightly to ensure firmness and reliability so as to avoid electric leakage After installation the power plug must be within reach so that the power plug can be unplugged in time in case of emergency Separate special outlets must be used and grounded reliably The cross sectional area of the copper 2 conductor in the wall connected with the outlet must be more than 4mm Do not lengthen the powe...

Page 5: ...r any outlet of the equipment otherwise electric shock or injury may be caused due to accidental contact between the above objects and moving parts In order to ensure the normal operation and ventilation and heat dissipation of the equipment the back left and right sides of the cabinet shall be at least 30cm away from the wall and the air inlet and air outlet must not be blocked by obstacles This ...

Page 6: ... cold and heat sources and strong electromagnetic interference which will affect the normal operation of the control system and directly damage the system in severe cases 4 2 Installation Site Figure N and E ends cannot be connected In order to operate the equipment normally and obtain the best performance level the installation site of the equipment shall meet the following requirements ò It cann...

Page 7: ...rt both ends of the handle cover into the corresponding grooves of the handle seat and then press it to close it 4 3 Preparation before use 4 4 First Power on When using the equipment for the first time please follow these steps 1 After the equipment is placed leveled and cleaned it shall stand for more than 24 hours and then power it on to ensure the normal operation of the equipment 2 Under no l...

Page 8: ...reached Electrical appliances without production license shall not be used inside the equipment Do not change the set temperature frequently in a short time otherwise the expected setting effect may not be achieved due to the large temperature inertia Ensure that there is a certain air circulation space around the cabinet when putting in items especially do not block the temperature sensor in the ...

Page 9: ...M Figure II GL RG 4M USB interface Door body Display panel Door lock Test hole Cabinet Caster Adjust feet Power switch Remote alarm interface Rs485 interface Door lock Test hole Door handle Cabinet Adjust feet Caster Door body USB interface Display panel ...

Page 10: ...8 Figure III Figure IV Caster Door handle Door body Cabinet Caster Door lock Adjust feet Steel wire shelf Test hole Printer optional Display panel power switch on the back USB interface ...

Page 11: ...he product consists of cabinet door glass door structure or foam door structure refrigeration system and control system Scope of application It is suitable for storing items in hospitals pharmacies epidemic prevention stations research institutions biopharmaceuticals and other units Figure V Steel wire shelf Door body Caster Cabinet Test hole Printer optional Display panel power switch on the back...

Page 12: ...ff When the printer is working the indicator is on f Door heating indicator When the door heating is turned off the door heating indicator is off When the door heating is turned on the door heating indicator is on g Refrigeration indicator If the compressor is in working condition the refrigeration indicator is on If the compressor is in a shutdown state the cooling indicator is off h Mute indicat...

Page 13: ...y for 3 seconds to import the data of the USB flash drive in 12 months 5 is a down key In parameter setting mode move to the previous parameter or decrease the parameter value For example when setting the set temperature reduce the set temperature When setting the parameter value long press the down button and the parameter will decrease rapidly 6 is a print key The system can keep 7 days of data f...

Page 14: ...0 5 0 Temperature setting 2 H 0 0 10 0 5 0 Set value of high temperature alarm set H When H 0 High temp alarm is disabled When the alarm is over high temp alarm set H1 will be displayed on the controller 3 L 0 0 10 0 4 0 Set value of low temperature alarm set L When L 0 Low temp alarm is disabled When the alarm is below low temp alarm set L1 will be displayed on the controller 4 Pt 0 240 min 20 Pr...

Page 15: ...arm L1 Low temperature alarm H2 Alarm for high ambient temperature H3 Condenser overheat alarm do Door opening alarm PF Power failure alarm bL Battery low alarm Er The recorder is not connected LoF Battery low alarm EE Communication failure 5 Alarm display ...

Page 16: ...t above the key is off e PRINT optional The system can keep 7 days of data for printing Press the print key to print the temperature within the set time period f In parameter setting mode reduce the parameter value For example when setting the set temperature reduce the set temperature When setting the parameter value long press the down button and the parameter will decrease rapidly g In paramete...

Page 17: ...d displays LoF the recorder is not started Meanwhile press and key for 3s and LoF disappears the recorder is started 2 Function setting of Type B control panel 1 Power on and turn on the power switch on the back of the cabinet and the equipment will enter the working state 2 User parameter settings Unlock In normal operation state simultaneously press and key for 3s and the digital tube displays t...

Page 18: ...is opened and closed again during the heating period the heating time will be updated again Mode 2 When the compressor runs Heater is on When the compressor stops the heater Will automatically off after one min Mode 3 When the humidity in the cabinet is more than 80 the door heating is on and when the humidity in the cabinet is moderately less than 60 the door heating is off Mode 4 Door heating is...

Page 19: ...ting paper has been installed at the factory When the paper is used up after a long time you may purchase the same paper for installation The installation steps are as follows Click the cylindrical button on the printer to open the cover of the printer Put the printing paper into the printer box and pull the paper head of the roll so that the paper slightly exposes from the upper port of the box c...

Page 20: ...and RS485 interface The remote alarm terminal is installed on the back of the cabinet and the alarm signal is output by the terminal The load bearing capacity of the terminal is 30V 2A DC Users can choose normally open normally closed and common terminals according to their own needs 18 RS 485 ...

Page 21: ...month and the power lock switch should be turned on so that the equipment can run for a period of time to complete charging and each charging time should be no less than 24 hours When the working power supply is interrupted the power lock switch of the equipment shall be turned off in time otherwise it will cause power loss in battery and even permanent damage to the battery in severe cases The ma...

Page 22: ...roduct may fail Please observe the operation of equipment in time during use If there is any abnormality please check and handle it according to the following table first If the abnormality can t be changed please inform our service center in time and we will serve you wholeheartedly to avoid losses Fan motor replacement time about 6 years Service life of the product 10 years Notes 20 After the eq...

Page 23: ...540X764 YC 130L 16 32 C N R134a 110g 220 230V 50 2 8 C 130 155 51 625X650X810 YC 315L 16 32 C N R600a 22g 220 230V 50 2 8 C 315 285 87 673X650X1762 YC 395L 16 32 C N R600a 25g 220 230V 50 2 8 C 395 290 95 673X650X1992 YC 395EL 16 32 C N R600a 25g 220 230V 50 2 8 C 395 250 89 673X650X1992 YC 400L 16 32 C N R290 65g 220 230V 50 2 8 C 400 500 116 645X700X2016 YC 525L 16 32 C N R290 75g 220 230V 50 2 ...

Page 24: ...mity Accessory package Key The door handle accessory applies to models YC 725L YC 1015L YC 1015EL and YC 1505L Keys apply to models YC 55L YC 55EL YC 75L YC 75EL YC 130L YC 315L YC 395L YC 395EL YC 400L YC 525L YC 725L and YC 1505L The specific packing list shall be subject to the physical objects received 1 1 1 2 Operation Manual ...

Page 25: ...EL YC 400L YC 525L YC 725L YC 1015L YC 1015EL YC 1505L Applicable to YC 130L The technical data and electrical schematic diagram shall be subject to the nameplate on the product and the electrical schematic diagram in case of any improvement of the product design We sincerely apologize for any inconvenience that might be caused ...

Reviews: