Functions 

 
Disinfection 

 

Flowmeter 

The  flowmeter  is  used  to  adjust  the  mixture  ratio  be-
tween  water  and  disinfectant.  It  is  located  at  the 
back  of  the  console.  It  can  only  be  calibrated  during 
a  disinfection  cycle  while  the  disinfectant  solution 
is  flowing.  The  table  next  to  it  should  help  you  to 
find  the  right  value.  Always  follow  the  instructions 
of  the  disinfectant  manufacturer.  The  Flowmeter  must 
be calibrated after each connection of water tubes and 
after changes in water pressure and after changing the 
disinfectant’s composition. During use of the tub, pay 
attention not to accidently alter its setting. 
 

Disinfection procedure

 

In  order  to  be  able  to  start  the  disinfection  proce-
dure  1)  there  must  be  disinfectant  inside  its  con-
tainer  and  2)  the  tub  must  be  almost  empty.  If  both 
requirements  are  met,  then  the  disinfection  button 
lights up. 
 

Note: 

If the disinfection container should be empty the dis-
infection button is flashing fast (~4x/sec.). 

 
To  start  the  disinfection  procedure  take  the 

red

  hand 

shower from its holder and press the 

disinfection

 but-

ton.  The  button  is  flashing  now.  Spray  all  internal 
tub surfaces with the disinfectant solution.  
 

Caution: 

Protect your skin and eyes from the disinfectant solu-
tion.  Follow  all  the  instructions  of  the  disinfectant 
manufacturer. 

 
Thereafter  press  the  button  again  and  put  the  hand 
shower back into its holder. 

After  the  necessary  contact  time  (according  to  the 
disinfectant’s  manufacturer’s  specifications)  the  tub 
must  be  rinsed  thoroughly.  To  do  so,  take  the 

white

 

hand  shower  from  its  holder  and  press  the 

shower

  but-

ton. Rinse all internal tub surfaces thoroughly. 

When  you  deactivate  the  shower  any  hydromassage  jets 
are activated briefly in order to remove the disinfec-
tant  solution  from  them.  Ensure  that  no  disinfectant 
solution residues are left behind inside the tub. 
 

Note: 

For  safety  reasons  the  disinfection  procedure  -  once 
started - cannot simply be aborted. 

 

 

 

During  the  whole  disinfection  procedure  most  other 
functions are disabled, especially the tub fill. 

Summary of Contents for GK Magic 1000

Page 1: ......

Page 2: ...and translation of both the whole document as well as of any parts thereof Congratulations on your decision to purchase a GK Bathtub You have acquired an Austrian quality prod uct which will considera...

Page 3: ...er 8 Setting the temperature 9 Temperature display tub fill shower 9 Underwater light 9 Sound Vision 10 Ultrasound 11 Ultrasound cleaning FAQ 11 13 Hydromassage 14 Shampoo Bath additive 14 Disinfectio...

Page 4: ...ce etc Never leave the patient unattended inside the bathtub Ensure that the equipment is thoroughly cleaned after every use When using disinfection protect your skin and eyes Heed all instructions on...

Page 5: ...4 3 2 1 8 11 10 5 6 7 Components Components 1 Disinfection 2 Lid 3 Shower 4 Control panel 5 Mixer 6 Shampoo Bath additive dispenser 7 Fill in spout 8 Overflow 9 Hydromassage jets 10 Drain 11 Grab han...

Page 6: ...5 Components Hydromassage jet Box for care and disinfection liquids USB port Flowmeter with con centration table...

Page 7: ...Sensor Activation card 6 Button Down 7 Button Up 8 Button Tub fill 9 Button Shower 10 Button Bath additive 11 Button Disinfection 12 Button Shampoo 13 Button Hydromassage Ultrasound 14 Button Underwa...

Page 8: ...oment and locked Only the sensor ACTIVATION CARD is illuminated with a green backlight The standby mode a prevents unauthorized use of the bathtub and b helps to save energy since due to the deactivat...

Page 9: ...n TUB FILL the tub starts filling The button starts flashing and the symbol TUB FILL TEMPERATURE lights up The water now flows into the bathtub until the maximum filling level has been reached or the...

Page 10: ...MPERATURE is illuminated additionally to the flashing button TUB FILL If the shower is active the symbol SHOWER TEMPERATURE is illuminated additionally to the flash ing button SHOWER If tub fill and t...

Page 11: ...on PLAY PAUSE during playback it is paused and the button starts to shine continu ously When pressing it again the playback contin ues In case the pause exceeds 30 seconds the player resets itself to...

Page 12: ...ination of the button stops and it can not be used until the next bathing cycle Note The ultrasound transducer is audible The closer any body parts are to the transducer plate the louder the side nois...

Page 13: ...and spread out through the bathing water What effect does ultrasound have regarding body cleaning Because of the ultrasound millions of tiny gas bubbles are created in the water which collapse again v...

Page 14: ...take unnecessary risks regarding the life of others May the ultrasound cleaning function be applied to people with pacemakers We have no information about any disruption of a pace maker s functionalit...

Page 15: ...massage hose coupling is installed then also the jet containing the coupling will be activated immediately without delay upon activation of the whole function Shampoo Bath additive Both of these prod...

Page 16: ...s infection button is flashing fast 4x sec To start the disinfection procedure take the red hand shower from its holder and press the disinfection but ton The button is flashing now Spray all internal...

Page 17: ...Always check the water temperature with your bare hand also when using the shower before bringing the water in contact with the patient To add bath additive to the water if installed available press t...

Page 18: ...y damage caused by unapproved unsuitable products will not be covered by the manufacturer s warranty Dark grey and transparent plastic parts The dark grey and transparent plastic parts may only be cle...

Page 19: ...h From time to time the hydromassage jet covers should be removed and be cleaned together with the inlet and outlet aper tures separately To do so unscrew the jet covers in counter clockwise direction...

Page 20: ...infection doesn t work and the button is flashing very fast about 4x sec even though the container is full The disinfection sensor is defective and has to be replaced All the functions are disabled ex...

Page 21: ...or other dam ages Caution Should you notice any damage to the bathtub either during maintenance or at any other time then the bathtub may not be used any further from that point onward until the damag...

Page 22: ...0 C Max relative humidity 80 not condensed at 20 C Atmospheric pressure range 500 hPa to 1060 hPa Transport and Storage Ambient temperature range 20 C to 60 C Relative humidity range 80 not condensed...

Page 23: ...on class I Degree of protection IPX4 Protection type Type B Lifting weight 400 kg Lifting height 400 mm Maximum length 2240 mm Maximum width 930 mm Maximum height 1250 1650 mm Time of guarantee 2 year...

Page 24: ...ents All product related information is not subject to change for technical production or other reasons Georg Kr mer Ges m b H Co KG A 4061 Pasching Neuhauserweg 14 Tel 43 0 7229 64450 Fax 43 0 7229 7...

Reviews: