Functions
Disinfection
Flowmeter
The flowmeter is used to adjust the mixture ratio be-
tween water and disinfectant. It is located at the
back of the console. It can only be calibrated during
a disinfection cycle while the disinfectant solution
is flowing. The table next to it should help you to
find the right value. Always follow the instructions
of the disinfectant manufacturer. The Flowmeter must
be calibrated after each connection of water tubes and
after changes in water pressure and after changing the
disinfectant’s composition. During use of the tub, pay
attention not to accidently alter its setting.
Disinfection procedure
In order to be able to start the disinfection proce-
dure 1) there must be disinfectant inside its con-
tainer and 2) the tub must be almost empty. If both
requirements are met, then the disinfection button
lights up.
Note:
If the disinfection container should be empty the dis-
infection button is flashing fast (~4x/sec.).
To start the disinfection procedure take the
red
hand
shower from its holder and press the
disinfection
but-
ton. The button is flashing now. Spray all internal
tub surfaces with the disinfectant solution.
Caution:
Protect your skin and eyes from the disinfectant solu-
tion. Follow all the instructions of the disinfectant
manufacturer.
Thereafter press the button again and put the hand
shower back into its holder.
After the necessary contact time (according to the
disinfectant’s manufacturer’s specifications) the tub
must be rinsed thoroughly. To do so, take the
white
hand shower from its holder and press the
shower
but-
ton. Rinse all internal tub surfaces thoroughly.
When you deactivate the shower any hydromassage jets
are activated briefly in order to remove the disinfec-
tant solution from them. Ensure that no disinfectant
solution residues are left behind inside the tub.
Note:
For safety reasons the disinfection procedure - once
started - cannot simply be aborted.
During the whole disinfection procedure most other
functions are disabled, especially the tub fill.