gk-medizinmechanik GK Magic 100 User Manual Download Page 9

 
Setting the temperature 

 
You  can  set  the  temperature  for  tub  fill  and  shower 
continuously.  To  do  that,  turn  the  knob  of  the  mixing 
valve  to  the  desired  position.  The  used  temperature 
will be shown in the display. 
 
 

Note: 

For  safety  reasons  the  temperature  is  mechanically 
limited  to  38  °C  (standard  mixer).  To  override  this 
security appliance, press the button on the knob. Thus 
it  is  possible  to  turn  the  knob  further  and  set  a 
higher temperature. 

 

Caution: 

When  setting  the  temperature  to  a  value  above  38  °C, 
set it back to 38 °C or lower as soon as possible! In 
the meantime, while the temperature is overridden, use 
extreme caution with the hot water. 

 
 

Temperature display tub fill & shower 

 
This display always shows the current tub fill respec-
tively shower temperature. 
 

Caution: 

Always check the water temperature with your bare hand 
(also when using the shower) before bringing the water 
in contact with the patient. 

 
 

Filling the tub 

 

Open the “TUB FILL VALVE” (in counter-clockwise direc-
tion)  to  fill  the  tub  with  water.  When  your  desired 
filling level is reached close the valve again to stop 
the water flow.

 

 
 
 

Filling the tub with control panel (if installed) 

 
If  the  knob  of  the  “TUB  FILL  VALVE”  is  not  present, 
then the tub can be filled by pressing the button “TUB 
FILL”.  Now  the  water  keeps  flowing  until  the  maximum 
filling  level  is  reached  or  the  button  is  pressed  a 
second time. If the water level is not high enough af-
ter  the  automatic  fill  stop,  you  can  press  and  hold 
the button “TUB FILL” for additional water flow. 
 

Note: 

If you want to use the door for  entering the bathtub, 
then  fill  the  bathtub  at  first  only  up  to  below  the 
doorstep. To do that when using the control panel, you 
must actively press the button at the right time. 

Functions 

Summary of Contents for GK Magic 100

Page 1: ......

Page 2: ...and translation of both the whole document as well as of any parts thereof Congratulations on your decision to purchase a GK Bathtub You have acquired an Austrian quality prod uct which will considera...

Page 3: ...ontrol panel 8 Shower 9 Shampoo Bath additive 9 Underwater light 9 Hydromassage 10 Sound Vision 10 Disinfection 11 12 Flowmeter 11 Manually controlled disinfection procedure model with disinfection va...

Page 4: ...ce etc Never leave the patient unattended inside the bathtub Ensure that the equipment is thoroughly cleaned after every use When using disinfection protect your skin and eyes Heed all instructions on...

Page 5: ...ure 5 Button Hydromassage lights up if configured 6 Button Down with printed arrow 7 Button Bath additive lights up if configured 8 Button Disinfection lights up if configured and needed Control panel...

Page 6: ...Adjustable foot 8 Foot switch 9 Control panel 10 Control panel for Sound Vision 11 Tub fill valve 12 Thermostat mixing valve 13 Fill in spout 14 Shampoo Bath additive dispenser 15 Disinfection valve...

Page 7: ...6 Components Hydromassage jets blue Box for care and disinfection liquids USB port Flowmeter with con centration table...

Page 8: ...y means of the foot switch It works with air pressure and is connected to the electronics through thin air filled plastic tubes The higher air pressure when stepping on the switch activates the corres...

Page 9: ...tion Always check the water temperature with your bare hand also when using the shower before bringing the water in contact with the patient Filling the tub Open the TUB FILL VALVE in counter clockwis...

Page 10: ...a red marking on its suc tion end the tube for the bath additive a blue mark ing at the alike spot If there is no liquid coming out of the system make sure that both boxes are full After installing a...

Page 11: ...2 mp3 wma is inserted into the USB port Now pressing the button PLAY PAUSE will play back all the recognized music files sorted by name Pressing the buttons NEXT and BACK switches to the next or previ...

Page 12: ...irectly on the hand shower and spray all internal tub surfaces with the disinfectant solution Caution Before every disinfection procedure ensure that there is enough disinfectant available in its cont...

Page 13: ...y all internal tub surfaces press the disinfection button again twice Thereby you are restarting the disinfection procedure and have an other 40 seconds available 5 After the 40 seconds are up and spr...

Page 14: ...it to a comfortable working height Once the desired filling level has been reached stop if necessary the inflowing water To add bath additive to the water if installed press the button BATH ADDITIVE f...

Page 15: ...p We especially recommend our special plastics intensive cleanser details on request For these parts of the tub only use very fine textured chamois leather or a microfiber cloth For the best care for...

Page 16: ...ed cover of the thermostatic mixing valve Also the cartridge could be defective and might have to be changed The display reads F0 or F1 The temperature sensor is defective and must be changed Very lit...

Page 17: ...d of the disinfection shower hose as well as of any externally visible connections Once a month check the bathtub for any cracks or other dam ages Caution Should you notice any damage to the bathtub e...

Page 18: ...0 C Max relative humidity 80 not condensed at 20 C Atmospheric pressure range 500 hPa to 1060 hPa Transport and Storage Ambient temperature range 20 C to 60 C Relative humidity range 80 not condensed...

Page 19: ...class I connection of protective earth conductor necessary Degree of protection IPX4 Protection type Type B Lifting weight 400 kg Lifting height 400 mm Maximum length 2240 mm Maximum width 930 mm Max...

Page 20: ...Tel 43 0 7229 64450 Fax 43 0 7229 70664 service gk medizinmechanik at www gk medizinmechanik at Typing spelling or transmission errors excepted for all contents All product related information is not...

Reviews: