ES
MANUAL DE USUARIO
STP
20
5. CALIBRATION
Turn on the crane scale first, then hold key
and press the key
until to watch 00000, in this moment set free the
keys and the display shows [
d *
]
CALIBRATION
STEP
OPERATION
DISPLAY
EXPLANATION
1
Press
to choose division value
(several times)
Divisin value: 1,2,5,10,20,50
Press the key
to confirm. Example: 2
d
*
2
P
*
Press
to choose decimal point
position (several times).
Decimal point optional: “0”,”0.0”,”0.00”,”0.000”,
Press
to confirm.
Example: 0.00
3
FULL
Setup F.S.
Input F.S.:Press
to choose the digit bit; press
to select the digit; press
to confirm the
input F.S..
Example
: 30000
4
noLoAd
Zero point calibration:
Press
when the stable light is
on.
Assure there’s no loading.
5
Load standard weights
6
AdLoAd
F.S. calibration: (Press
directly
if the value to be input is same as F.S.
when stable light is on)
Input weight of standard weights loaded: Press
to choose the digit bit; press
to select the
digit. Until the input value is same as the loaded
weight and the digit bit is at the highest bit, press
to confirm when the stable light is on.
6. ERROR INDICATION
1
Err 1
AD value is too small when calibration.
2
Err 2
Zero point is out of range when calibration.
3
Err 3
Zero point is out of range when power on
4
Err 5
Weight input is 0 when full scale calibration
5
bAt-Lo
Indicator out of battery, please charge.