After introducing the parameters LF2, the indicator will show the last configuration saved. All the
steps to follow must be completed, if they are not done the indicator will continue with the previous
configuration.
Proceed with the calibration of weight after LF2.
LINEAL CALIBRATION
LF 3
DISPLAY
DESCRIPTION AND SEQUENCE OF USE
LF3
•Press the Key
to start or the key
To exit the menu and the scale will start again automa-
tically
W0
•Press the Key
to calibrate the zero.
( press the key
o exit from the calibration and go back to the menu LF3)
W1
• Place 1/3 of the maximum weight on the plate and press
to calibrate 2/3 th the capacity. (press
the key
to exit from the calibration and go back to the menu LF3)
W2
•Place 2/3 of the maximum weight on the plate and press
to calibrate 2/3 th the capacity.
(press the key
to exit from the calibration and go back to the menu LF3)
W3
•Place the maximum weight on the plate and press the key
to calibrate the full capacity. (press
the key
to exit from the calibration and go back to the menu LF3)
SPEED OF THE AD CONVERTER
LF 4
SPEED 1
standard speed 15Hz.
SPEED 2
high speed 30Hz.
SPEED 3
low speed 7.5Hz
*This function stays blocked when UF-5 is in mode HOLD 1.
*The value of factory is 1
DISPLAY
DESCRIPTION AND SEQUENCE OF USE
LF4
•Press the key
to start or the key
to exit from the menu and the scale is going to begin
again automatically.