EN
USER’S GUIDE
FFN
19
7 - 6 . R S 2 3 2 O U T P U T U F - 6
232 0 - RS232 disable
232 4 - Stable output – Format 2
232 1 - Stable output – Format 1
232 5 - Stream output – Format 2
232 2 - Stream output – Format 1
232 6 - Manual output – Format 2
232 3 - Manual output – Format 1
7 - 7 . B A U D R AT E
b 1200
1200
b 9600
9600
b 2400
2400
b 19200
19200
b 4800
4800
b 38400
38400
1. Press the key
to access to auto-power configuration.
2. Press
to select the baud rate.
3. Press the key
to confirm.
7 - 8 . C O M M U N I C AT I O N P R O T O C O L
UART signal of EIA-RS232 C
Format:
5. Serial output:
1200/2400/4800/9600/
19200/34800 BPS
6. Data bits: 8 bits
7. Parity bits: None
8. Stop bits: 1 bit
Start bit
Data bits
Stop bits
F O R M AT 1 ( 2 3 2 1 – 3 ) :
HEAD1 (2 bytes)
HEAD 2 (2 bytes)
OL - Overload
ST - Stable
NT - Net weight
US - Unstable
GS - Gross weight
F I X E D 1 8 BY T E S A S C I I
( k g g t l b )
1
2
1
1
2
1
1
2
3
4
5
6
7
8
1
2
1
2
HEAD1
HEAD2
DATA
UNIT
CR LF
F I X E D 2 1 BY T E S A S C I I
( t l . T l b o z )
1
2
1
1
2
1
1
2
3
4
5
6
7
8
9
1
2
3
4
1
2
HEAD1
HEAD2
DATA
UNIT
CR LF
F I X E D 1 9 BY T E S A S C I I
( p c s )
1
2
1
1
2
1
1
2
3
4
5
6
7
8
1
2
3
1
2
HEAD1
HEAD2
DATA
UNIT
CR LF
1. E0.876 kg Stable net weight:
S
T
,
N
T
,
+
0
0
0
.
8
7
6
k
g
0D
0A
2. Example -1.568 lb unstable gross weight:
U
S
,
G
S
,
-
0
0
1
.
5
6
8
l
b
0D
0A
3. Example -20. 5.40 lb oz unstable gross weight:
S
T
,
G
S
,
-
1
0
.
0
5
.
4
0
l
b
o
z
0D
0A
4. E1000 pcs stable net weight:
S
T
,
N
T
,
+
0
0
0
1
0
0
0
P
c
s
0D
0A