39
EN
USER’S MANUAL
ABD
LF3
•Press the Key
to start or the key
to exit the menu and the scale will start again auto-
matically.
W0
•Press the Key
to calibrate the zero.
( press the key
to exit from the calibration and go back to the menu LF3)
W1
• Place 1/3 of the maximum weight on the plate and press
to calibrate 2/3 th the capacity.
(press the key
to exit from the calibration and go back to the menu LF3).
W2
•Place 2/3 of the maximum weight on the plate and press
to calibrate 2/3 th the capacity.
(press the key
to exit from the calibration and go back to the menu LF3).
W3
•Place the maximum weight on the plate and press the key
to calibrate the full capacity. (press
the key
to exit from the calibration and go back to the menu LF3).
W4
•Press the key
to complete the lineal calibration. (press the key
to exit from the cali-
bration and go back to the menu LF3)
LF3
•Use the keys
,
and then the key
to continue with other adjustments and press the
key
to exit from the menu and the scale will begin again automatically.
VELOCIDAD DEL CONVERTIDOR AD
LF 4
SPEED 1
standard speed 15Hz.
SPEED 2
high speed 30Hz.
SPEED 3
low speed 7.5Hz
*This function stays blocked when UF-5 is in mode HOLD 1.
*The value of factory is 1
DISPLAY
DESCRIPTION AND SEQUENCE OF USE
LF4
•Press the key
to start or the key
to exit from the menu and the scale is going to begin
again automatically.
SPEED 1
•Use the key
and then the key
to select the speed of the AD converter ( press the key
to exit from the configuration and go back to the menu LF4)
LF 4
•Use the keys
,
and then
. To continue with the other adjustment and press the key
to exit from the menu and the scale will start again automatically.
BLIND
LF 5