37
EN
USER’S MANUAL
ABD
CALIBRATION OF THE WEIGHT
LF 1
DISPLAY
DESCRIPCIÓN y SECUENCIA DE USO
The calibration can be done with any weight, but the weight ca not be inferior to 1/100 of the
maximum capacity and it must also never be exceeded.
•Press the key
to start the calibration of zero ( press
to exit the calibration and go back
to the menu LF1)
LF 1
CAL Z
150.00 kg
•Use
,
and
,
0 ~ 9
and then
pto introduce the weight with which the cali-
bration will be done. ( press the key
to exit the calibration and go back to the menu LF1)
150.00 kg
•Place the required weight on the scale as it is indicated in the display.
150.00 kg
•Once everything is steady, press the key
to calibrate it ( press the key ESC to exit the calibra-
tion and go back to the menu LF1).
THE CALIBRATION IS GOING TO FINISH AND THE SCALE WILL GO VACK TO THE WEIGHING MODE AUTOMATICALLY.
CONFIGURATION
LF 2
DISPLAY
DESCRIPTION AND SEQUENCE OF USE
FIRST STEP
•DISPLAY OF THE INTERNAL COUNTING
262144
SECOND STEP
•POSSIBLE VALUES OF THE PARAMETERS
A: Metric system
0:NO
1: kg
2:T
3:g
1 0 0 0 0 1
B: American system
0:NO
1: lb
2:lb oz
C: other unities
0:NO
1: TW kg
2:HK kg
3:VISS
D: PCS
0:OFF
1: ON
A B C D E F
E: double range
0:OFF
1: multi interval
2:multi range
F: units of calibration
1:use metric
unities
2:use american unities
lb oz cannot be selected as unities of calibration.
The scale won’t let us continue up to the next step if there is a mistake during the programming.
THIRD STEP
•Use
,
and
,
0 ~ 9
and then press
to introduce THE MAXIMUM CAPACITY.
000000kg
FOURTH STEP
•Use the Keys
,
to change the position of the DECIMAL POINT.
dP 0.0
d 0.00
d 0.000
d 0.0000
0.00000
FIFTH STEP
•Use
to select the DIVISION:
diV 01, diV 02, diV 05, diV10, diV 20, diV 50
diV 01