Guidance and manufacture’s declaration – electromagnetic immunity – for spirometer
Guidance and manufacture’s declaration – electromagnetic immunity
The spirometer (SP10) is intended for use in the electromagnetic environment specified below. The customer or
the user of SP10 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If
floor are covered with synthetic material, the rela-
tive humidity should be at least 30%.
Power frequency
(50/60Hz) magnetic
field IEC61000-4-8
3A/m
3A/m
Power frequency magnetic fields Should be at lev-
els characteristic of a typical location in a typical
commercial or hospital environment.