![background image](http://html.mh-extra.com/html/gima/primaled-so/primaled-so_operators-manual_2228147014.webp)
Operator
’
s Manual
MO080-IT
12/09/2017
Rev.3
Page 14 of 20
7
Technical properties
Technical properties
PRIMALED/ PRIMALED-FLEX
Illumination Ec at 50cm distance ± 10% [Lux]
95,000 / 110,000
Colour temperature (±5%) [K]
4,000 / 4,500
Colour rendering index Ra [-]
95
Max irradiation [W/m2]
294
350
Irradiation / Illumination [mW/m2lx]
3.09
3.18
Max irradiation in UV [W/m2]
0.002
Focalization from grip
No
Power connection details
Primary alternate voltage [Volt ac]
100-240
Frequency [Hz]
50/60
Power input [VA]
22 (100V) – 30 (240V)
Light source
N°9 Led
Duration of LED diode light source [hr]
(this figure can vary according to power peaks and operating
frequency)
50,000
Light intensity control [%]
25 –100
General data
Colour
RAL 9003
Directive
2007/47/EC
Standards
IEC 60601-2-41
Classification of Medical Device
Class I
Essential performance
Distribution of minimum and adequate lighting (luminous flux emitted by the ME equipment
does not vary by more than 20% during use and the colour temperature and the colour
rendering index are stable and are within the range 3000K-6700K and 85-100, respectively)
Limitation of energy in the operating field (UV-irradiance for wavelengths below 400 nm
does not exceed 10 W/m
2
and the total irradiance Ee in the lighted area does not exceed 1000
W/m
2
at a distance of 500 mm).
IP Classification
IP20
Operating conditions
Continuous operation
Mains power voltage insulation means
Integrated power plug
Integrated fuses
T1AH 250V, 5x20
Handpiece steam sterilization
121°C 1.3 bar from 25 to 30 minutes
134°C 2.3 bar for 4 minutes.
Dimensions
Diameter of lamp body [cm]
19.5
Light emission surface [cm
2
]
42
63
Lamp weight [Kg]
12
Markings
In conformity with Directive 93/42/EEC (and
2007/47/EC)
All technical light measurements are to be deemed with a tolerance of ±6% for metrological and manufacturing
reasons
Summary of Contents for PRIMALED SO
Page 15: ...Operator s Manual MO080 IT 12 09 2017 Rev 3 Page 15 of 20 8 Wiring diagram ...
Page 21: ...Drawing code Rev Data 052 1 2 3 ...
Page 22: ...Drawing code Rev Data 053 1 2 3 ...
Page 23: ...Drawing code Rev Data 054 2 1 3 4 ...
Page 25: ...Drawing code Rev Data 056 2 1 3 4 5 ...
Page 26: ...Drawing code Rev Data 057 3 1 2 ...
Page 27: ...Drawing code Rev Data 058 2 1 3 4 ...
Page 28: ...Drawing code Rev Data 1 2 7 3 4 5 6 153 ...
Page 29: ...Drawing code Rev Data 190 1 2 3 4 5 6 8 7 ...
Page 30: ...Drawing code Rev Data 191 1 2 4 5 3 6 7 ...
Page 31: ......
Page 32: ......