Guidance and Declaration - Electromagnetic Immunity
The Blood pressure monitor is intended for use in the specified
electromagnetic environment. The customer or the user of the Blood
pressure monitor should assure that it is used in such an
environment as described below.
Immunit
y test
IEC606
01 test
level
Complian
ce level
Electromagnetic
environment - guidance
Conduc
ed RF
IEC6100
0-4-6
3 Vrms
150 kHz
to
80 MHz
outside
ISM
bands
3
Vrms
150 kHz
to
80 MHz
outside
ISM
bands
Portable
and
mobile
RF
communications
equipment
should be used no closer to
any part of the system,
including cables, than the
recommended
separation
distance calculated from the
equation appropriate for the
frequency of the transmitter.
Recommended
separation
distances:
P
d
2
.
1
Radiate
d RF
IEC6100
0-4-3
3
V/m
80 MHz
to
2.5 GHz
3 V/m 80
MHz to
2.5 GHz
Recommended
separation
distances:
80
MHz
~
800
MHz:
P
d
2
.
1
800MHz-2.5GHz:
P
d
3
.
2