3. Test Record: this menu to review the records, user can print the record,
delete record, and clear all record.
4.1. Review all records by pressing “
←
” or “
→
” icon.
4.2. Print: print out the test record, also can print out the photo after show the photo.
4.3. Delete: to delete the test record, need input user password
4.4. Exit: return to previous menu.
4.5 Delete all record: need input user password.
4. Sensor Test:
Press “Sensor Test” icon and enter the sensor test
screen. Press “Auto” icon, then blow
alcohol
breath
(about 0.05%BAC) into chamber through
the blow tube for about six seconds and observe the
sensor sensitivity & waveform as attached.
If the
cure is a straight line, it means the sensor
already lost the sensitivity and need to change a
new sensor.