8.1 Battery Maintenance
Keep the both sides of coin cell clean.
- Low temperature may decrease the performance of coin cell, and low
battery indicator may appear early. In such case, please put coin cell
into pocket for warm before use, thus bring it back to normal condi-
tion.
- Do not let any conductive metal (such as tweezers) contact both sides
of coin cell simultaneously to avoid short circuit.
- Charge the coin cell for 8~10 hours each time; ambient temperature
should be 5°C~40°C.
- If the coin cell is full after charging, but its performance decreases
apparently, it means the coin cell is exhausted, please change a new
one.
9. TROUBLESHOOTING
Trouble
Possible reason
Solution
The SpO
2
and Pulse Rate
display instable.
1. The finger is not
placed inside enough.
2. The finger is shaking
or the patient is
moving.
1. Place the finger
properly and try again.
2. Let the patient keep
calm.
The device
can not turn on.
1. The coin cell is
drained or almost
drained.
2. The coin cell is not
inserted properly.
3. The device’s
malfunction.
1. Charge the coin cell.
2. Reinstall coin cell.
3. Please contact the
local service center.
The indicator
light is off
suddenly.
1. The device will
power off automatically
when it gets no signal
for 8 secs.
2. The coin cell is
almost drained.
1. Normal.
2. Charge the coin cell.