- 34256 Rear valve (intake valve)

- 34259 Non-Rebreathing valve

Instruction for use 

Product Description: 

Gima 

Manual Resuscitator is designed for use as an adjunct to artificial respiration

 

and cardiopulmonary 

resuscitation. The resuscitator can be used to ventilate the 

apnoeic patient and to augment ventilation and 

oxygen delivery to the spontaneously

 

breathing patient. The designs are also different according to adult, 

child and infant by

 using dif

ferent compressing frequency. They also come with different sizes to meet all

 

patients’ need for oxygen.

 

Specification 

Connectors: 

patient port

         

      15 mm l.D.(F)/22 mm O.D.(M) 

SILICONE BAG INLET 

     23 mm |.D.(M) 

reservoir valve        

 23 mm |.D.(F) (to silicone bag inlet) 

      25 mm O.D.(M) 

(to oxygen reservoir)

 

supplemental gas

 

inlet       

     6 mm O.D.(M)   

Shelf Life   

      5 

years 

 

Dead Space:

       

  

Pressure Relief Valve:

 

non-rebreathing valve    7ml 

 Child and infant     

      40 ± 5cmH2O 

 adult mask  

   150ml 

 Adult     

 60 ± 10cmH2O 

child mask   

 95ml 

infant mask   

 28ml 

Summary of Contents for 34259

Page 1: ...ario sia un adulto un bambino o un neonato usando una diversa frequenza di compressione Sono inoltre disponibili varie misure per soddisfare ogni diversa esigenza di ossigeno nel paziente Caratteristi...

Page 2: ...0 Neonato B 40 40 40 I risultati sono stati ottenuti nelle seguenti condizioni Adulto Vr 600mL Compliance 0 02L cmH2O Resistenza 20cm H20 L s Bambino A Vr 70mL Compliance 0 01L cmH2O Resistenza 20cm H...

Page 3: ...o maggiore di 30 kg Avvertenze 1 Il prodotto deve essere conservato in un luogo asciutto e pulito e mantenuto sigillato 2 L operatore deve essere addestrato e conoscere le tecniche di rianimazione car...

Page 4: ...Dispositivo medico conforme alla Direttiva 93 42 CEE Leggere le istruzioni per l uso Leggere e seguire attentamente le istruzioni per l uso Conservare al riparo dalla luce solare Data di fabbricazion...

Page 5: ...ent The designs are also different according to adult child and infant by using different compressing frequency They also come with different sizes to meet all patients need for oxygen Speci cation Co...

Page 6: ...ditions Adult Vr 600mL Compliance 0 02L cmH2O Resistance 20cm H20 L s Child A Vr 70mL Compliance 0 01L cmH2O Resistance 20cm H20 L s Child B Vr 300mL Compliance 0 01L cmH2O Resistance 20cm H20 L s Inf...

Page 7: ...ld be placed in a dry and clean place and sealed conditions 2 The operator should be trained and know the skill in CPR resuscitation Precautions The patient should be closely monitored by quali ed med...

Page 8: ...aly Made in Taiwan BMIU O2 04 Medical Device complies with Directive 93 42 EEC Consult instructions for use Caution read instructions warnings carefully Keep away from sunlight Date of manufacture Exp...

Reviews: