background image

 

Bag 

volume 

Stroke 

volume 

Reservoir 

volume 

Suitable body

 

weight

 

Adult Model 

Child model 

Infant model 

1600ml 

500ml 

280ml 

700ml 

300ml 

150ml 

2500ml 

2500ml 

500ml 

> 30 kg 

7-30 kg 

<7 kg 

 

Stroke volume of 1350ml can be achieved using 

two

 hands 

Minimum Cycle Rate 

                                                                  

Oxygen Concentration:

 

Adult         20 breaths/min                                                           

with reservoir 

          99% 

Child          20 breaths/min                                                           

without reservoir 

   45% (adult and child models) 

Infant         40 breaths/min                                                                                               90% (infant model) 

 

The performance characteri

stics for Manual Resuscitat

ors will

 

vary from user to user depending on a 

variety of factors:

 

ambient temperature, patient lung compliance, ventilatory

 

frequency, size of operator's 

hands. 

 

Maximum Ventilatory Rates

 

 

-18°C (0°F) 

Cycle Rate

 

22°C (72°F) 

Cycle Rate

 

22°C (72°F) 

Cycle Rate

 

Adult 

20 

20 

20 

Child A 

30 

30 

30 

Child B 

20 

20 

20 

Infant A

 

60 

60 

60 

Infant B

 

40 

40 

40 

 

The results were obtained under the following conditions:

 

Adult:

 Vr-

600mL, Compliance 0.02L

/

cmH2O, Resistance 20cm H20/L/s

 

Child 

A:

 Vr-

70mL, Compliance 0.01

L/cmH2

O, Resistance 20cm H20/

L/s 

Child B:

 Vr-

300mL, Compliance 0.01

L/cmH2

O, Resistance 20cm H20/L/s

 

Infant A: Vr

-

20mL, Compliance 0.001L

/cmH2

O, Resistance 400cm H20/L/s

 

Infant B:

 Vr-

70mL, Compliance O.01

L/

cmH2O, Resistance

 20cm H20/L/s 

 

The correct ventilation frequency may vary, please follow

 

current ventilation frequencies as recommended 

by major specialist associations

Range of delivery pressure

 

Adult: 6O

±10cm H2O maximum 

(can be overridden by operator)

 

Child and infant

: 40

±5 cm H2O maximum 

(can be overridden by operator)

 

 

Indication for us

e: 

Gima s

ingle use manual resuscitators are indicated to temporarily augment ventilation in patients.

 

Gima Silicone manual Resuscitators are 

intended to provide emergency ventilation o

r ventilation during 

patient transport, with air or air enriched with oxygen. The infant model is intended for patients weighing 

up to

 

10 kg, the child model is intended for patients weighing between 10 and 30 kg, and the adult model is 

Summary of Contents for 34259

Page 1: ...ario sia un adulto un bambino o un neonato usando una diversa frequenza di compressione Sono inoltre disponibili varie misure per soddisfare ogni diversa esigenza di ossigeno nel paziente Caratteristi...

Page 2: ...0 Neonato B 40 40 40 I risultati sono stati ottenuti nelle seguenti condizioni Adulto Vr 600mL Compliance 0 02L cmH2O Resistenza 20cm H20 L s Bambino A Vr 70mL Compliance 0 01L cmH2O Resistenza 20cm H...

Page 3: ...o maggiore di 30 kg Avvertenze 1 Il prodotto deve essere conservato in un luogo asciutto e pulito e mantenuto sigillato 2 L operatore deve essere addestrato e conoscere le tecniche di rianimazione car...

Page 4: ...Dispositivo medico conforme alla Direttiva 93 42 CEE Leggere le istruzioni per l uso Leggere e seguire attentamente le istruzioni per l uso Conservare al riparo dalla luce solare Data di fabbricazion...

Page 5: ...ent The designs are also different according to adult child and infant by using different compressing frequency They also come with different sizes to meet all patients need for oxygen Speci cation Co...

Page 6: ...ditions Adult Vr 600mL Compliance 0 02L cmH2O Resistance 20cm H20 L s Child A Vr 70mL Compliance 0 01L cmH2O Resistance 20cm H20 L s Child B Vr 300mL Compliance 0 01L cmH2O Resistance 20cm H20 L s Inf...

Page 7: ...ld be placed in a dry and clean place and sealed conditions 2 The operator should be trained and know the skill in CPR resuscitation Precautions The patient should be closely monitored by quali ed med...

Page 8: ...aly Made in Taiwan BMIU O2 04 Medical Device complies with Directive 93 42 EEC Consult instructions for use Caution read instructions warnings carefully Keep away from sunlight Date of manufacture Exp...

Reviews: