7
—
Attention: The ECG medical gel coating should be separated, and the touching chest electrode may
result in short circuit.
6.2.2 Limb electrode
Place the electrode on the hand and leg, clean the positions around to which limb electrodes are to
be attached with alcohol, then lay some ECG medical gel on the cleaned skin.
Attention: Tighten the knob after inserting the mainframe patient cable pin in the patient cable
socket.
6.2.3 Electrode connection definitions and color code:
Electrode
Location
Electrode
Code
Line
Color
Electrode
Color
Socket
Number
Right arm
R
Black
Red
9
Left arm
L
Black
Yellow
10
Left leg
F
Black
Green
11
Right leg
RF
Black
Black
14
Chest 1
V1/C1
White
Red
12
Chest 2
V2/C2
White
Yellow
1
Chest 3
V3/C3
White
Green
2
Chest 4
V4/C4
White
Brown
3
Chest 5
V5/C5
White
Black
4
Chest 6
V6/C6
White
Purple
5