![Gima 32916 Manual Download Page 52](http://html1.mh-extra.com/html/gima/32916/32916_manual_2227740052.webp)
52
SPECIFICATIONS
WARNING: No modification of this equipment is allowed.
Power supply
3.7V 1000mAH Built-in rechargeable
li-polymer battery, 6V
1A AC Adaptor
Display mode
Digital LCD display
V.A.= 86.1mm(L) x24mm(W)
Measurement mode
Oscillographic testing mode
Measurement range
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Accuracy
Pressure:
5°C-40°C within ±3mmHg (0.4kPa)
Pulse value: ±5%
Normal working condition
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour partial
pressure greater than 50 hPa
An atmospheric pressure range of : 700 hPa to 1060 hPa
Storage & transportation
condition
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
Measurement perimeter
of the arm
About 22cm-42cm
Weight
Approx.284g
External dimensions
Approx.130.9mm×73mm×29.4mm
Attachment
AC Adaptor, user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against
ingress of water
IP22: The first number 2: Protected against solid foreign
objects of 12,5mm Ф and greater. The second number:
Protected against vertically falling water drops when
enclosure titled up to 15º. Vertically falling drops shall
have no harmful effects when the enclosure is titled at any
angle up to 15º on either side of the vertical
Software version
A01
Device classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor charged Mode:
Class II ME Equipment
ENGLISH