background image

17

PERIODICAL MAINTENANCE CHECKS

The VEGA SUCTION ASPIRATOR does not need maintenance or lubrication. 

It is, however, necessary to inspect the unit before each use. With regard to training, given the information 

contained in the user manual and since it is easy to understand the said device, it doesn’t appear to be nec-

essary. 

Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been 

damaged during previous use. Connect the cable to electrical network and turn the switch on. 

Close the aspirator outlet with your finger and with suction regulator at maximum check that the vacuum 

indicators reaches at least -75kPa (-0.75 bar). Rotate the knob from right to left. The vacuum indicator should 

go down -25kPa (-0.25 bar). 

Check that no loud noises are present. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and 

situated in the plug protects the instrument. For fuses replacing, always the type and the range. 

Before changing the fuse, disconnect the plug from the power supply socket.

If the overfill security system it’s activated, don’t proceede with the liquid aspiration. 

If the overfill security system doesn’t work there are two cases:

1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter 

who avoid the liquid penetration inside the device. 

2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, 

in this case return the device to GIMA technical service. 

Gima S.p.A. will provide upon request electric diagrams, components list, descriptions, setting in-

structions and any other information that can help the technical assistance staff for product repair.

 BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, 

PLEASE CONTACT GIMA TECHNICAL SERVICE. GIMA DOES NOT GIVE GUARANTEE IF INSTRU-

MENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED. 

ENGLISH

Fault type

Cause

Solution

1. The suction unit doesn’t 

work

Cable is damaged External power 

source failure

Replace the cable Check the external power 

source

2. No aspiration

Jar Cap not screwed on correctly

Unscrew the cap, and re-screw it

3. No aspiration

Lid seal not in its seat

Unscrew the cap and insert the seal properly 

in its seat

4. The Vacuum power on the 

patient side is either very low 

or absent

•  Vacuum regulator set 

  to minimum

•  Protection filter blocked or damaged

•  Connection tubes blocked, kinked 

or disconnected

•  Shut-off valve blocked or 

 damaged

•  Pump motor damaged

•  Turn the vacuum regulator clockwise and 

check the value of the vacuum on the gauge

•  Replace the filter

•  Replace or reconnect the tubes, check the jar 

connections 

•  Empty the jar, or disconnect the tube from the 

jar and unblock the shut-off valve. The unit 

twill only work in the upright position 

•  Refer to authorised service personnel

5. The float doesn’t close

If the cap has been washed, ensure 

that the float is not partially detached

Fit the float into it’s place

6. The float doesn’t close

The float it’s covered by dirty material Unscrewed the cap, leave the and put in on 

autoclave

7. Low suction

Foam inside the jar

Fill the jar to 1/3 full of ordinary water

8. No aspiration due to flow 

leakage of mucus  

Filter blocked  

Replace filter  

Faults 1 - 2 - 3 - 4 - 5 - 6 

- 7 - 8

None of the procedures have 

achieved the desired results

Contact GIMA customer service

Summary of Contents for 28210

Page 1: ...HIRURGISCHER ABSAUGER VEGA ASPIRADOR QUIR RGICO VEGA APARELHO DE SUC O CIR RGICO VEGA M28210 M Rev 7 12 20 PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly...

Page 2: ...rtment Do not attempt to make the device work before it has been thorough ly checked by qualified personnel and or the GIMA technical service department Position the device on stable and flat surfaces...

Page 3: ...nce they contain small parts that may be swallowed 14 Do not leave the device unattended in places accessible to children and or persons not in full possession of mental facultiesas they may strangle...

Page 4: ...on pump is used on patients in unknown pathological situations and where an assess ment of indirect contamination is not possible The filter is not manufactured to be decontaminated dismantled and or...

Page 5: ...of local laws and regulations The jug and lid can be further disinfected using a common disinfectant strictly following the instructions and dilution values provided by the manufacturer At the end of...

Page 6: ...nents list descriptions setting in structions and any other information that can help the technical assistance staff for product repair BEFORE EVERY CHECKING OPERATION IN CASE OF ANOMALIES OR BAD FUNC...

Page 7: ...the presence of liquids or other visible contamination secretions In the presence of liquids or other visible contamination immediately replace the medical device due to the risk of an insufficient v...

Page 8: ...d cables sold by the appliance and system manufacturer as spare parts can lead to an increase in emissions or in a decrease of the immunity of the device or system The following tables supply informat...

Page 9: ...ains power quality should be that of a typi cal commercial environment or hospital If the user of the VEGA SUCTION UNIT request that the appliance operates continuously the use of a continuity unit is...

Page 10: ...es such as changing the device s direction or positioning may be necessary b The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V m Recommended separation distance...

Page 11: ...EEE disposal Class II applied Serial number Temperature limit Fuse Atmospheric Pressure limit IP21 Covering Protection rate Alternating current ON Hz Mains frequency OFF Humidity limit SYMBOLS Disposa...

Reviews: